LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Contact PCB (A07) Replacement
359
8
Contact PCB (A07) Replacement
Contact PCB replacement consists of the following procedures:
•
Removing the Contact PCB (A07) (p. 359)
•
Installing the Contact PCB (A07) (p. 359)
Removing the Contact PCB (A07)
) (from outside the rear case) (see
1.
Lay the device face down on a static-free, non-abrasive surface.
2.
) between the battery wells. Lift away the battery
retainer. Discard the screws.
3.
Lift the Contact PCB away for rear case far enough to access and
disconnect the power/contact PCB cable (
) (W05) at
J42.
Installing the Contact PCB (A07)
To install the contact PCB (
) on the outside of the rear case:
1.
Lay the device face down on a static-free, non-abrasive surface.
2.
Connect the contact PCB to the power/contact PCB cable (W05) at J42.
3.
Install the contact PCB by sliding the assembly straight down tracks in rear case.
4.
Install the two new screws (
) into the battery retainer (
) between the battery wells; torque to
10.0 in-lb using P2 bit.
contact
PCB
battery
retainer
screws
Figure 8.122—Contact PCB screw and part locations