LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
ECG Connector Cable (W07) Replacement
329
8
Removing the ECG Connector Cable (W07)
Refer to
To remove the ECG connector cable (
) from the rear case (refer to
Inside Rear Case Diagrams (p. 228)
)
:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Disconnect the ECG cable from the J6 connector on the system PCB (A01).
3.
Remove the parameter bezel as described in
Removing the Parameter Bezel (p. 311)
.
4.
Remove the four screws (
) securing the support bracket (
) that holds the ECG connector cable
). Remove the connector, seal and screws. Discard the seal and screws.