LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Biphasic Module (A22)/Inductive Resistor (A14) Replacement
299
8
Removing the Biphasic Module (A22)/Inductive Resistor (A14)
To remove the biphasic module and/or the inductive resistor from the rear case (refer to
Inside Rear Case Diagrams (p. 228)
)
:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
Therapy (A04) PCB Assembly Replacement (p. 232)
.
3.
Remove the screw (
) securing the clear, plastic, high-voltage
shield (
to the interconnect bracket (A17). Discard the screw.
4.
Cut the two cable ties on the right side of the interconnect bracket.
5.
Disconnect the three biphasic module spade terminal connectors from the
interconnect bracket (A17 - pin 3, pin 6, and pin 9). Needle nose pliers
may be used to assist in disconnection.
cover screw
Figure 8.94—Interconnect bracket screw locations
DANGER
SHOCK HAZARD
Lethal voltages may be present even without
operator action. Always discharge the energy storage capacitor
prior to servicing. See the service manual “
” for detailed instructions.