Instructions for Use | Technical Specifications
Specifications
90
Monitored Parameter Accuracy
Airway pressure
±(2 cmH
2
O + 4% of actual)
Tidal volume
± (4 ml + 15% of actual) for volumes ≥ 35 ml
± 10 ml for volumes < 35 ml
FiO
2
±(2.5% FiO
2
+ 2.5% of actual reading) within a 24
-hour 2-point calibration
period, or a change in altitude
EtCO
2
(with Mainstream CO
2
sensor)
±(
volume fraction of 0.43% + 8% of gas level
)
SpO
2
and pulse rate
See the sensor manufacturer’s operating instruction sheet.
Operational range of the Patient Circuit and Circuit accessories
Inspiratory/expiratory resistance
•
Pediatric/adult circuit (20
-
22 mm): less than 5 cmH
2
O at 30 l/min
•
Pediatric/adult circuit (19 mm): less than 5 cmH
2
O at 15 l/min
•
Infant/pediatric
circuit (14
-
16 mm): less than 5 cmH
2
O at 15 l/min
•
Infant/pediatric circuit (9
-
13 mm): less than 5 cmH
2
O at 2.5 l/min
Compliance
Up to 4 ml/cmH
2
O
Wireless
Bluetooth
Operating frequency range
2402-2480 MHz
Channel bandwidth
1 MHz/2 MHz
Maximum output power
12.0 dBm
Modulation
GFSK, Pi/4 DQPSK, 8DQPSK
Near-
field communication (NFC)
Operating frequency
13.56 MHz
Receiving section bandwidth
1.4 MHz
Maximum output power
23.0 dBm
Modulation
ASK, OOK
Wi-Fi
Operating frequency range
2402-2480 MHz
Channel bandwidth
20 MHz/40MHz
Maximum output power
17 dBm
Modulation
DSSS, OFDM, DBPSK, DQPSK, CCK, 16-QAM
Security
Wpa2
All flows and volumes are expressed in BTPS
.
Measurement Uncertainty for Control and Performance Specifications
The stated tolerances account for the measurement uncertainty of the test equipment used to verify performance:
Pressure: ± 0.75% cmH
2
O
Tidal Volume: ±2 ml
Oxygen: ±1% FiO
2