Instructions for Use | Device Setup
Installing Filters
31
4.
Device Setup
Setup Overview
To set up Trilogy Evo, follow the steps shown below. See the accompanying section for instructions.
Step
Section
Page
1.
Place the device.
Placement
31
2.
Connect AC power.
Connecting AC Power
31
3.
Install filters.
Installing Filters
32
4.
Connect a circuit.
Connecting a Circuit
32
5.
Add oxygen (optional).
Adding Oxygen
35
6.
Start the device.
Starting Trilogy Evo Universal 36
Placement
Place Trilogy Evo on a stable, flat, hard surface. Air must flow freely. Do not block the air vents with items such as bedding or curtains. Do not place
Trilogy Evo near any heating or cooling equipment or air supplies such as forced air vents, radiators, or air conditioners. Ensure the USB and
detachable battery panel doors remain closed when not in use.
If the device has been stored outside the normal operating temperature stated in “Technical Specifications,” ensure the device reaches operating
temperature before connecting power.
See the “EMC Information” chapter for guidance on possible electromagnetic interference.
Connecting AC Power
Use the AC cord provided to connect AC power before pressing the power button. Verify Trilogy Evo Universal is using AC power, indicated by the
green LED light next to the power button.
To use other power sources, such as the detachable battery or an external battery, see the “Power Management” chapter.
Installing Filters
CBRN Filter
To install the CBRN filter:
1.
Ensure the CBRN filter adapter gasket is securely seated.
2.
Place the filter adapter onto the bayonet mount and twist clockwise to lock in place.
3.
Screw the CBRN filter onto the filter adapter.