Operation
6-11
Figure 6-3: PPV initial setup
Start
Make initial settings:
EPAP
: 4 cmH
2
O
O
2
: Current setting or per prescription
Max P
: 25 cmH
2
O
Max V
: 1000 to 1500 mL
PPV %
: 80 to 100%
Max E
: 5 cmH
2
O/L
Max R
: 2 cmH
2
O/L/s
HIP
: > Max P
Other settings: Per usual protocol
Yes
No
Is the patient using
accessory muscles to ac-
tively stop inspiration?
Was the Max V or Max P
limit reached?
Has the mask leak suddenly
increased?
No
No
No
Increase Max E by
2 cmH
2
O/L
Yes
Yes
Yes
No
Does the patient say the air
is coming too fast?
Is the patient using
accessory muscles to ac-
tively stop inspiration?
Was the Max V or Max P
limit reached?
Has the mask leak suddenly
increased?
No
No
Increase Max R by
1 cmH
2
O/L/s
Reduce Max R by 1 cmH
2
O/L/s for
patient comfort
Yes
Yes
Yes
Yes
No
End
Titrate PPV % downward as tolerated
Does the patient say they
are getting too much air,
pressure, or volume?
Reduce Max E
by 2 cmH
2
O/L for patient comfort
Max E
titration
Repeat Max E titration as needed
Max R
titration
Содержание Respironics V60
Страница 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Страница 2: ......
Страница 8: ...viii ...
Страница 28: ...2 10 Symbols ...
Страница 44: ...3 16 General information ...
Страница 60: ...4 16 Principles of operation ...
Страница 74: ...5 14 Setting up the ventilator for use ...
Страница 102: ...6 28 Operation ...
Страница 110: ...7 8 High flow therapy ...
Страница 114: ...8 4 Patient monitoring ...
Страница 132: ...9 18 Alarms messages and troubleshooting ...
Страница 142: ...10 10 Care and maintenance ...
Страница 162: ...A 10 First time installation ...
Страница 182: ...D 10 Regulatory compliance ...
Страница 198: ...E 16 Diagnostic mode ...
Страница 210: ...Index Index 6 ...
Страница 211: ......