11-8
Technical specifications
Electrical
specifications
Accessory
requirements
NOTE:
Any part that is added to the patient circuit changes the compliance
and resistance of the ventilator breathing system. To achieve the
ventilator performance specified in this manual, the ventilator
breathing system must meet the specifications listed in Table 11-11.
Table 11-10: Electrical specifications
Parameter
Specification
AC voltage
100 to 240 VAC
AC frequency
50/60 Hz
AC power
300 VA
Battery
PN 1076374: 14.4 V, 11.0 Ah, 163 Wh
Maximum system current draw: 11 A
Charge voltage: +16.9 V maximum
Operating time: 360 minutes (6 hours) under normal
conditions
Table 11-11: Accessory Requirements
Parameter
Specification
Compliance
For volume accuracy requirements, the patient circuit
compliance should be 0.98 mL/cmH
2
O.
For all other performance requirements except
volume accuracy, the maximum compliance of the
circuit can be up to 2.8 mL/cmH
2
O.
Resistance
Maximum resistance of the breathing circuit and at-
tachments: 2.9 cmH
2
O at 60 L/min.
Bacteria filter
Dead Space: 66 ml
Bacterial/Viral Filter Efficiency: >99.99%
Resistance (@ 0.5 L/s): 0.7 cmH
2
O (hPa)/L/s
Male Connector: 15 mm I.D./22 mm O.D.
Female Connector: 22 mm I.D.
Содержание Respironics V60
Страница 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Страница 2: ......
Страница 8: ...viii ...
Страница 28: ...2 10 Symbols ...
Страница 44: ...3 16 General information ...
Страница 60: ...4 16 Principles of operation ...
Страница 74: ...5 14 Setting up the ventilator for use ...
Страница 102: ...6 28 Operation ...
Страница 110: ...7 8 High flow therapy ...
Страница 114: ...8 4 Patient monitoring ...
Страница 132: ...9 18 Alarms messages and troubleshooting ...
Страница 142: ...10 10 Care and maintenance ...
Страница 162: ...A 10 First time installation ...
Страница 182: ...D 10 Regulatory compliance ...
Страница 198: ...E 16 Diagnostic mode ...
Страница 210: ...Index Index 6 ...
Страница 211: ......