15
2
.
2.4
.
2
.
Amplitude·modulated (AM) H
F
·
s
i
gn
al generator
Internal
-
Pre liminary adjustment according to chap.
2.2.4. 1 . .
- Press button AM/ 1 kHz. The HF-signal is modu lated in amp litude by 30 % at an internal frequency of 1 k Hz.
-
For indications on checking and adjusting, see the ind ividua l
check ing
and adjusting procedures for the
object to be measured or chapter 2.3 . .
External
- Preliminary adjustment according to chap. 2.2.4.1
.
.
- Press button AM/EXT.
-
Supply the modulation voltage to connector
LF IN; maximum 16
V
peak-peak,
m aximum
d.c.
level
±12 V.
- Externa l modulation frequency 20 Hz ... 20 kHz.
- Adjust the modulation depth by the ampl itude of the modulation voltage; voltage necessary 0.2 V / 10 %
ampl itude·modu lation.
For indications on checking and adjusting, see the individual checking and adjusting procedures for the
object to be measured or chapter 2.3.
2.2.4.3. Frequency·modulated (FM) H F-signal generator
Internal
- Press buttons 75 - 1 10 or 10.3 - 11.1 MHz.
The selected range scale is indicated by a LED.
- Set the pointer at the required frequency by potentiometer FREQUENCY.
Not marked frequency va lues should be interpolated between two mar ks.
-
If necessary, calibrate the frequency at the nearest ca l i bration-mar k
according
to sub·para
2.2.4.5
.
.
- Press button FM/1 k Hz; the HF·signa l
is
modulated in
frequency
by
1
kHz and
a
frequency sweep of
about
25 kHz.
- Apply the signal, available at BNC connector H F OUT, across a cabl e to the object to be measured.
(see accessories, chapter 1.3.).
- Adjust the required output voltage by means of potentiometer HF A MPLITUDE.
For indications on checking and adjusting, see the individual checking and adjusting procedures for the
object to be measured or chapter 2_3.
Note
When using harmonics the frequency sweep is multiplied by the order-number of the harmonics.
External
- Pre liminary adjustment according to chap. 2.2.4.3 . .
- Press button F M/EXT.
- Supply the modulation vo ltage to BNC connector LF IN; m aximum input voltage 70 V peak-peak.
- Adjust the external modulation frequency; 20 Hz ... 60 kHz.
-
Adjust the modulation sweep with the externa l
voltage; necessary voltage 0.2 V17.5 kHz
sweep
(maximum
sweep 75 k Hz).
- If necessary, supply a multiplex-signal to BNC connector LF IN.
For indications on checking and adjusting, see the individual check ing and adjusting procedures for the
object to be measured or chapter 2.3.
2.2.4. 4. Wobble generator
- Press range-button 75 ... 110 or 10.3 ... 1 1.1 MHz.
The selected range-scale is indicated by a LED.
- Set the pointer to the required frequency by potentiometer FREQUENCY.
Not marked frequency va lues should be interpolated.
- Press button WOB.
- Connect BNC-connector 1 k Hz/SWEEP OUT to the X-input of an oscilloscope.
- Adjust the X-amplitude of the osc i l loscope.
- Apply the wobbu lated signal across a cable to the object to be tested.
Содержание PM5324
Страница 1: ...HF Generator 100kHz 110MHz PM5324 9452 053 240 1 9499 450 05511 730801 1 01 ...
Страница 15: ...14 Fig 4 Front view Fig 5 Rear view ...
Страница 43: ...Fig 33 Front view with item numbers Fig 34 Bottom view ...
Страница 44: ...44 Fig 35 Top view Fig 36 Left side view ...
Страница 45: ...45 0 0 0 0 1 ji I I MAB225 Fig 37 Scale dri ve n ...