Manual Report Filters
PageWriter 100 M1772A Cardiograph
2-9
2
0DQXDO5HSRUW)LOWHUV
The PageWriter 100 allows you to choose between two sets of filters
for Manual ECG reports:
l
The 0.05-40 Hz filters (0.05 Hz Baseline Wander and 40 Hz
Noise) are enabled when the filter light is on. It delivers the
least amount of noise.
l
The 0.05-150 Hz filters (0.05 Hz Baseline Wander and 150 Hz
Noise) are enabled only when the Filter light is off. It delivers
the highest fidelity signal, but unless ECG signal conditions are
excellent, it can record noise.
Table 2-3 shows when each filter is enabled.
7DEOH0DQXDO)LOWHU6HWWLQJV
Default (Filter light on)
Filters Off (Filter light off)
0.05-40 Hz
0.05-150 Hz
a
a. Conforms to AAMI specifications