Auto Report Filters
PageWriter 100 M1772A Cardiograph
2-5
2
7DEOH/HDGV2II/DEHOV
$XWR5HSRUW)LOWHUV
The PageWriter 100 allows you to choose between two sets of filters
for Auto ECG reports:
l
The 0.15-40 Hz filters (0.15 Hz Baseline Wander and 40 Hz
Noise) are enabled when the Filter light is on. It delivers the
least amount or noise.
l
The 0.15-150 Hz filters (0.15 Hz Baseline Wander and 150 Hz
Noise) are enabled only when the Filter light is off. It delivers
the highest fidelity signal, but unless ECG signal conditions are
excellent, it can record noise.
Table 2-2 shows when each filter is enabled.
7DEOH$XWR)LOWHU6HWWLQJV
Designator (AHA/IEC)
Meaning
RL/N
Right leg electrode not connected, or only right leg
electrode is connected and all other limb electrodes
are not connected.
RA/R
Right arm electrode is not connected.
LA/L
Left arm electrode is not connectd.
LL/F
Left leg elecrode is not connected.
V1 ... V6/
C1 ... C 6
One or more chest electrodes are not connected. For
example, V2 means the V2 electrode is not con-
nected.
Default (Filter light on)
Filters Off (Filter light off)
0.15-40 Hz
0.15-150 Hz
a
a. Conforms to AAMI specifications
Содержание M1772A
Страница 1: ...I N S T R U C T I O N S F O R U S E C a r d i o g r a p h PageWriter 100 M1772A ...
Страница 2: ......
Страница 3: ...PageWriter 100 M1772A Cardiograph 3DJH ULWHU 0 DUGLRJUDSK ...
Страница 10: ...Contents 4 ...
Страница 22: ...AC and DC Battery Operation 1 12 Getting Acquainted ...
Страница 32: ...Understanding the Printed Report 2 10 Recording an ECG 8QGHUVWDQGLQJ WKH 3ULQWHG 5HSRUW LJXUH XWR 5HSRUW ...
Страница 38: ...Choosing a Report Format 2 16 Recording an ECG LJXUH 0DQXDO HDG ...
Страница 44: ...Identifying ECG Problems 3 6 Troubleshooting ...
Страница 60: ...Calling for Service 4 16 Maintaining the Cardiograph Spain Tel 34 91 631 31 00 ...
Страница 76: ...Glossary Glossary 4 ...
Страница 80: ...Index 4 ...
Страница 81: ......
Страница 82: ...3DUW 1XPEHU 0 3ULQWHG LQ 86 GLWLRQ ...