Draft - 1 Aug 08
Telemetry Functions at the Information Center
Telemetry Functions at the Information Center & TeleMon
9-7
SpO
2
Enabled
Enable/disable the SpO2
measurement at the Information
Center or Telemon.
enable
disable
enable
SpO
2
Mode
Determine the transceiver SpO
2
behavior.
Note—
Pulse Rate and Pleth Wave
are not available in
Spot Check
.
Spot Check
- Provides
manual measurements so the
clinician can check as
needed. Measurement
initiated by plugging the
SpO
2
cable into the
transceiver or by selecting
the Spot Check SpO
2
icon in
the Patient Window. No
pulse or successful/failed
measurement tones are
generated when
measurement is initiated at
Information Center.
Continuous
- Sends an SpO
2
parameter value to the
Information Center every
second. If selected, Pulse
Rate and Pleth Wave may
also be sent.
Spot Check
Patient-Configurable Settings in Telemetry Setup
Control
Function
Setting Choices
Factory
Default
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...