Network Design
4-10
Site Planning and Preparation
Printers
HP LaserJet printers
can be network connected to the network for shared printing
capability.
A summary of these design capabilities and limitations of Patient Care Network
components is given in
Information
Center (Local
Database)
This section includes independent Information Center designs only, that is Information
Centers that are
not
connected to a IntelliVue Clinical Network.
Equipment
Capability
Requirements must be matched to the following Philips Patient Care Network equipment
capabilities.
•
Information Center
capability can:
– display data from up to
16 patient monitors
– annunciate
22 alarms
– store up to
150 alarm records
– store up to
96 hours o
f
patient data for:
—
waves
—
ST segments
—
trends
—
events
•
each
System Communications Controller
can manage data for up to
6 Information
Centers
and from up to
24 patient monitors
.
•
Patient monitors
for both systems can be
hardwired or telemetry
.
•
Telemetry mainframes
can support
8 channels of telemetry
monitoring.
These design capabilities and limitations are summarized in Table 4-2, Philips Patient Care
Network Component Capabilities and Limitations.
Table 4-1. Release E.01 System Capabilities
Description of Limit
M3150 Local
Database
M3169 Small
Database
M3154 Central
Database
Maximum Number of Information Centers per system
1
3
8
Maximum Number of Clients per system
0
3
8
Maximum Number of Database Servers per system
0
1
10
Maximum Number of beds per Information Center
16
4
16
Maximum Number of monitored beds per system
16
48
128
Maximum Number of networked printers
2
4
8
Table 4-2. Philips Patient Care Network Component Capabilities and Limitations
Component
Device
Capabilities
Limitations
Patient Monitors
Hardwired monitor
1 per patient
Telemetry monitors
1 per patient
Telemetry Mainframe
8 monitors per Mainframe
See
Note
below
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...