M3500B/M5500B Service Manual
3
-
5
Diagnosing Internal Failures
3
Diagnosing Internal Failures
The following steps will help you isolate an internal failure to a particular
subassembly.
Run Self Tests
To run the Self Tests, perform the following steps:
1. Enter Diagnostic Mode. See "Accessing Diagnostic Mode" on page 3-6.
2. Run the Extended Self Test. See "Extended Self Test" on page 2-17.
3. Run the individual Diagnostic Tests that are applicable to the problem.
Diagnostic Tests are described starting with "User Interface Tests" on
page 2-19.
Use the Troubleshooting Tables
Use the Tables provided starting on page 3-8 to isolate the problem based on
factors such as:
z
Error codes reported in the System Log (Table 3-2, page 3-11).
z
Extended Self Test failures (Table 3-6, page 3-18).
z
Operational problems (starting with Table 3-7 on page 3-19).
Configuration and Diagnostic Modes
Some of the Troubleshooting procedures look for responses from the unit that
are dependent on that unit’s configuration. Similarly, many procedures
require that the unit be in Diagnostic Mode. Follow the instructions given
below to enter these modes.
Accessing Configuration Mode
These instructions describe briefly how to enter Configuration Mode. See the
Instructions for Use
for details on configuration settings and what effect they
have.
1. Make sure the power is off.
2. If you intend to save the configuration to a Data Card (or load the config-
uration from a Data Card), insert the Data Card now.
3. Press softkeys 4 and 5 at the same time, and hold them down while turn-
ing the power on. See Figure 3-1 for softkey numbering.
z
To select a configuration, press the
and
softkeys to move up
or down the list until the desired
Settings
item is highlighted. Then
press the
softkey to access those settings.
z
To print out a strip with all the current configuration choices, select
Print All Settings
and press
.
ENTER
ENTER
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...