Guidance and manufacturer’s declaration-
electromagnetic immunity
The Philips wearable biosensor-G5 Solution is intended for use in the
electromagnetic environment specified below. The customer or the user
of the Philips wearable biosensor-G5 Solution should assure that it is
used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electrostatic discharge
(ESD) IEC 61000-4-2
± 8kV contact
± 15kV air
± 8kV contact
± 15kV air
Electromagnetic Environment Guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Immunity Test
IEC 60601 Test Level
Compliance Level
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
30 A/m
30 A/m
Electromagnetic Environment Guidance
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.