• Beaker of water (or water bath)
Verifying the Biopsy Guide Alignment
WARNING
If the needle enters from the unexpected side of the display or cannot be seen, verify that the
biopsy guide is correctly mounted on the transducer and that the orientation of the
transducer is correct. If the needle is still not following the expected path along the guideline,
do not use the biopsy guide. Contact your Philips representative.
1. Attach the biopsy guide. Although some transducers require the use of a second transducer
cover for biopsy procedures, a second transducer cover is unnecessary for this alignment
verification.
2. Connect the transducer to the system, and select the appropriate applications and preset.
3. Set the system depth for the procedure to be performed.
4. Display the biopsy guideline.
5. Without changing the position of the biopsy depth cursor, note the default depth shown at
the bottom of the display.
6. Immerse the transducer no more than 6 mm (0.25 in) into the water bath.
Biopsy Guides
Biopsy Guide Alignment
250
EPIQ 7 User Manual 4535 617 25341
Содержание epiq 7
Страница 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Страница 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Страница 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Страница 100: ...Brake Steering Lock Pedal System Overview System Components 100 EPIQ 7 User Manual 4535 617 25341 ...
Страница 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Страница 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Страница 198: ...Intraoperative Transducers Leakage Current Testing for Intraoperative Transducers 198 EPIQ 7 User Manual 4535 617 25341 ...
Страница 244: ...Endocavity Transducers Biopsy with Endocavity Transducers 244 EPIQ 7 User Manual 4535 617 25341 ...
Страница 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...