- 53 -
Heater plate
Max Temperature: 68°C (154°F)
Humidity
Humidity
min
Output:
≥
12 mg H
2
O/L
Measured at expected leak across therapy pressures , 17 °C– 35°C,
15% RH for all compatible tubes.
Maximum delivered
gas temperature
<43°C (109.4°F)
Declared dual-number noise emissions values (in accordance with ISO 4871)
Tube Size
Sound Pressure
Level
Uncertainty
Sound Power
Level
Uncertainty
15 and 22 (mm)
tubing type
27 dB(A)
2 dB(A)
35 dB(A)
2 dB(A)
12 (mm) tubing
type
27 dB(A)
2 dB(A)
35 dB(A)
2 dB(A)
Pressure accuracy
Pressure Increments: 4-20 cmH
2
O (in 0.5 cmH
2
O increments)
Maximum static and dynamic pressure accuracy, according to ISO 80601-2-70:2015:
Tube Type
Static
Dynamic
4
to 20 cmH
2
O
15 and 22 (mm) tubing type
± 0.5 cmH
2
O
± 1.0 cmH
2
O
12 (mm) tubing type
± 1.0 cmH
2
O
± 2.0 cmH
2
O
Static pressure accuracy has a measurement uncertainty of 1.31%.
Dynamic pressure accuracy has a measurement uncertainty of 2.70%.