- The effectiveness of this blood pressure moni­

tor has not been established in pregnant
(including pre-eclamptic) women.

- Common arrhythmias (such as atrial or ventric­

ular premature beats or atrial fibrillation) and
peripheral artery disease / arteriosclerosis can
affect the performance (accuracy) of this blood
pressure monitor. Please consult your doctor
how to best use this blood pressure monitor if
you suffer from any of these conditions.

- Only use this device for its intended purpose as

described in this user manual.

- Do not confuse self-monitoring with self-diag­

nosis. This device allows you to monitor your
blood pressure. Do not begin or end medical
treatment based on the measurement results.
Always consult your doctor for treatment
advice.

- Do not take any therapeutic measures on the

basis of a self-measurement. Never change
prescribed medication without consulting your
doctor. Consult your doctor if you have any
questions about your blood pressure.

- If you are taking medication, consult your

physician to determine the most appropriate
time to measure your blood pressure. 

- This device is not intended for use on extremi­

ties other than the wrist or for functions other
than obtaining a blood pressure measurement.

- If the cuff pressure exceeds 40kPa (300mmHg),

the unit will deflate automatically. If the cuff
does not deflate when pressures exceeds
40kPa (300mmHg), detach the cuff from the
wrist and press the 'on' button to stop inflation.

- Do not attach the cuff on the same arm on

which other monitoring medical electrical
equipment is attached simultaneously, because
this could cause temporary loss of function of
those simultaneously-used monitoring medical
electrical equipment.

- Never attach the cuff on injured skin, an injured

arm or an arm under medical treatment as this
can cause further injury.

- Do not use the device in case of existing poly­

ester or nylon material allergies.

- This device is not washable. Never immerse the

device in water and do not rinse it under the
tap.

- This device is not suitable for continuous moni­

toring during medical emergencies or opera­
tions.

- This device cannot be used with HF (High Fre­

quency) surgical equipment at the same time.

- Never use compressed air, scouring pads, abra­

sive cleaning agents or aggressive liquids such
as petrol or acetone to clean the device.

- If the battery can no longer be recharged or

when the device does not function properly
anymore (see 'Specifications'), please contact
the Philips Consumer Care center in your coun­
try.

- Keep the device away from fire and heat

sources, as the battery can overheat, causing
fire or bursting. The battery could explode,
causing injury or death. 

- After charging, remove the micro USB plug from

the device and remove the USB plug from USB
portal.

- The equipment is not AP/APG equipment and

is not suitable for use in the presence of a flam­
mable anesthetic mixture with air, with oxygen
or nitrous oxide.

- To avoid measurement errors, do not use the

device near strong electrical or magnetic fields,
for example magnets, radio transmitters,
microwave ovens.

- To avoid measurement errors, do not use the

device near a strong electromagnetic field radi­
ated interference signal or electrical fast tran­
sient/burst signal.

- Use this device under the right environmental

conditions as indicated in this user manual. If
not, this could affect the performance, lifetime
of the device and measurement results. 

- Use a soft cloth to clean the whole unit. Do not

use any abrasive or volatile cleaners.

- Only use the micro USB cable supplied to

charge the device. 

- The device does not need to be calibrated for

two years of normal use. After two years the
measurements may become less accurate.

- If you have any problems with this device, such

as setting up, malfunction, maintaining or using,
please contact Philips Consumer Care.

- Do not open, disassemble or repair the device

yourself. Modification of the device voids your
warranty.

- Report to Philips Consumer Care if any unex­

pected operation or events occur.

- Dispose of accessories, detachable parts, and

the ME equipment according to the local guide­
lines.

- Do not attempt to replace your blood pressure

monitors battery. It is built-in and not change­
able.

- Avoid charging your blood pressure monitor in

extremely high or low temperatures (see 'Speci­
fications').

- Do not clean the blood pressure monitor when

it is being charged. Always unplug the charger
first before cleaning the blood pressure moni­
tor.

Compliance with standards

- The device meets the relevant standards for this

type of Class IIa electrical medical equipment
for home use.

- This Philips device complies with all applicable

standards and regulations regarding exposure
to electromagnetic fields and complies with EN
60601-1-2.

Display

Sym

bol

Description Explanation

Systolic

blood pres­

sure

Maximum blood pressure.

Also see section Systolic

and diastolic pressure

Содержание DL8765

Страница 1: ...lijke Philips N V All rights reserved Philips Consumer Lifestyle BV Tussendiepen 4 9206AD Drachten Netherlands Fax 31 0 512594316 4222 100 4515 2 10 2015 1 2 2 3 4 DL8765 5 6 7 8 9 10 1 2 cm 0 4 0 8 11 12 13 14 15 16 17 18 19 20 75 recycled paper 75 papier recyclé ...

Страница 2: ...ctly and to protect you and others from injury Below you find the meaning of the warning signs and symbols on the label and in the user manual Symbol for follow instructions for use This symbol means that the part of the device that comes into physical contact with the user also known as the applied part is of type BF Body Floating accord ing to IEC 60601 1 The applied part is the cuff Symbol for ...

Страница 3: ... transfu sion or an arterio venous shunt A V shunt is present The temporary interference to blood flow by the blood pressure measurement could result in injury If you had a mastectomy breast amputation do not use this blood pressure monitor on the arm on the side of the mastectomy The inflat ing cuff can lead to pain trauma and further injury in the arm on the side of the mastectomy Consult your d...

Страница 4: ...tery can no longer be recharged or when the device does not function properly anymore see Specifications please contact the Philips Consumer Care center in your coun try Keep the device away from fire and heat sources as the battery can overheat causing fire or bursting The battery could explode causing injury or death After charging remove the micro USB plug from the device and remove the USB plu...

Страница 5: ...charging indications Battery symbol Battery charging indication Battery charging half full Battery charging almost full Battery fully charged Using the blood pressure monitor This tubeless device uses the oscillometric method to measure blood pressure and heart rate Before every measurement the unit establishes a zero point equivalent to the atmospheric pres sure Then it starts inflating the cuff ...

Страница 6: ...re monitor is equipped with Blue tooth Smart You can receive your personal health data on a mobile device that is equipped with the Bluetooth Smart function Download the Philips HealthSuite health app from the App store or Google Play Use the search term Philips Health Suite health app The app is available for iOS 8 0 and Android 4 4 Note You can only use the Philips HealthSuite health app to comm...

Страница 7: ... automatically selects the previous user To change the user profile press the user ID button and select a different user Fig 14 Make sure the correct user is selected so the measurement data is properly transmitted and stored Also a guest user can be selected The guest user is for performing a measurement on other persons without a user profile in the app Measurements performed when using the gues...

Страница 8: ... for the separate collection of electrical and electronic products and rechargeable batteries Correct disposal helps prevent negative consequences for the environment and human health Removing the rechargeable battery Warning This procedure is irreversible You cannot use the device anymore after this pro cedure Note We strongly advise you to take your product to an official collection point or a P...

Страница 9: ...ked position Fig 7 The light of the display dims and a battery symbol Lo is showing The battery is low Charge the battery see Charging The display shows Err Communica tion error Check if the app is on and try data transmission again The display shows E3 The cuff is not properly secured Refasten the cuff wait for 3 minutes and then measure again The display shows E10 or E11 The device detected moti...

Страница 10: ...c environments EMC Guidance The Blood Pressure Monitor needs special pre cautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the accompanying documents Wireless communications equipment such as wireless home network devices mobile phones cordless telephones and their base stations walkie talkies can affect this equipment and should be ...

Страница 11: ...e quency 50 60 Hz magnet ic field IEC 61000 4 8 3A m 3A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commer cial or hospital environment Note UT is the AC mains voltage prior to applica tion of the test level Guidance and manufacturer s declara tion electromagnetic immunity for ME equipment and ME systems that are not life supporting Guida...

Страница 12: ...ation distances between portable and mobile RF communications equip ment and the device The device is intended for use in an electromag netic environment in which radiated RF distur bances are controlled The customer or the user of the device can help prevent electromagnetic inter ference by maintaining a minimum distance between portable and mobile RF communications equipment transmittters and th...

Отзывы: