![Philips AVENT SCF391/71 Скачать руководство пользователя страница 24](http://html.mh-extra.com/html/philips/avent-scf391-71/avent-scf391-71_manual_286650024.webp)
Single breast pump
Double breast pump
Cycle speed
Stimulation: 105-120 cycles/min, depending on vacuum level
Expression: 53-85 cycles/min, depending on vacuum level
Motor unit rated input
5 V d.c. / 1.1 A
5 V d.c. / 1.8 A
Mains adapter rated input External mains adapter 100-240 V a.c. / 50/60 Hz
Mains adapter rated
output
5 V d.c. / 1.1 A
5 V d.c. / 1.8 A
Adapter type number
S009AHz050yyyy The letters "yyyy" represent the output current from
0100 (1000
mA) to 0180 (1800
mA), increments in steps of 100
mA. The
letter "z" denotes type of plug, it can be: U, V, B, S, C, A, D, K, T, I, E, F.
Power port
Micro-USB
Operating conditions
Temperature: 5
°C to 40
°C (41
°F to 104
°F)
Relative humidity: 15 to 90
% (non-condensing)
Storage and
transportation conditions
Temperature: -25
°C to 70
°C (-13
°F to 158
°F)
Relative humidity: up to 90
% (non-condensing)
Operating pressure
700-1060
hPa (<3000
m/9842 ft. altitude)
Net weight
230
g
310
g
External dimensions
Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
Device classification
Applicable classifications:
Degree of protection against electric shock;
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specified as medical electrical
system according to IEC 60601-1.
Materials
ABS, silicone (motor unit)
Polypropylene, silicone (other parts)
Service life
500
hours
Mode of operation
Continuous operation
Ingress protection
IP22 (motor unit), IP20 (adapter)
Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic
compatibility (EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying
tables. Only use this Philips Avent single/double electric breast pump Advanced in combination with
the adapter supplied.
24
English
Содержание AVENT SCF391/71
Страница 2: ......
Страница 3: ...D C B A B1 B2 B3 B4 B5 B6 C5 C3 C4 C2 C1 A1 A2 A3 A4 A5 D7 D1 D3 D5 D4 D6 D8 D2 ...
Страница 4: ......
Страница 28: ...Symbol Explanation Indicates the operating pressure range 700 to 1060 hPa 28 English ...