Emissions
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The device does not use RF energy for its intended
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in locations in
residential environments and in establishments
directly connected to a low voltage power supply
network which supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
CompliesClass A Category
Voltage fluctuations / flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Plus is intended for use in the electromagnetic environment listed
below. The user of the device should ensure that it is used in such an environment to prevent improper operation.
Immunity test
Immunity test level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Floors can be a source of static electricity. When the air
is very dry (<20 % RH) the user is advised to prefer to
use the device in a room with concrete or wooden floors
(instead of using it in a room with a synthetic floor).
Radiated RF EM fields
IEC 61000-4-3
10 V/m80 MHz – 2.7 GHz80 %
AM at 1 kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
385 (27), 450 (28), 710 (9), 745
(9), 780 (9), 810 (28), 870 (28),
930 (28), 1720 (28), 1845 (28),
1970 (28), 2450 (28), 5240 (9),
5500 (9), 5785 (9) MHz (V/m)
The device is immune against any RF wireless
communication device such as smart phone, Wi-Fi
or Bluetooth device at a distance of 30 cm or more.
Avoid a closer distance as this may result in improper
operation.
Rated power
frequencymagnetic fields
IEC 61000-4-8
30 A/m50 Hz or 60 Hz
Electrical fast transient/ burst
IEC 61000-4-4
±2 kV100 kHz repetition
frequency
Mains power quality expected of a typical commercial
or residential environment.
Surges
IEC 61000-4-5
±1 kV Line-to-line±2 kV Line-
to-ground
Mains power quality expected of a typical commercial
or residential environment.
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3 Vrms, before modulation is
applied
0.15 MHz – 80 MHz
6 Vrms, before modulation is
applied, in ISM and amateur
radio bands between 0.15 MHz
and 80 MHz
80 % AM at 1 kHz
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0 % UT; 0.5 cycle at 0°, 45°,
90°, 135°, 180°, 225°, 270° and
315°
0 % UT; 1 cycle and 70 % UT;
25/30 cycles single phase
at 0°
0 % UT; 250/300 cycle
Mains power quality expected of a typical commercial
or residential environment. At power interruptions the
breast pump will switch off.