15 Monitoring Uterine Activity Internally
135
Troubleshooting
Internal (IUP) Monitoring
Problem
Possible Causes
Solutions
No change in pressure during
contraction.
Dry environment or possible extra-
ovular placement of sensor tip.
Please refer to catheter Instructions for Use.
Only pressure peaks can be
seen (baseline not visible).
Zero adjustment is incorrect.
Zero the system.
Trace is a straight line.
Transducer is defective.
Remove and touch the catheter. If the trace
does not show up and down movements, use a
new transducer.
Catheter blocked.
Please refer to catheter Instructions for Use.
Trace is superimposed with
noise.
End of catheter is in the uterine wall. Please refer to catheter Instructions for Use.
IUP EQUIP MALF
INOP is displayed.
See “Patient Alarms and INOPs” on page 73.
IUP UNPLUGGED
INOP is displayed.
Содержание Avalon FM20
Страница 2: ......
Страница 8: ...8 ...
Страница 56: ...3 Basic Operation 56 ...
Страница 72: ...5 Alarms 72 ...
Страница 82: ...6 Patient Alarms and INOPs 82 ...
Страница 86: ...7 Admitting and Discharging 86 ...
Страница 90: ...8 Non Stress Test Timer 90 ...
Страница 120: ...11 Monitoring Twin FHRs 120 ...
Страница 126: ...12 Monitoring Triple FHRs 126 ...
Страница 136: ...15 Monitoring Uterine Activity Internally 136 ...
Страница 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Страница 154: ...18 Monitoring SpO2 154 ...
Страница 164: ...20 Printing the ECG Waveform 164 ...
Страница 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Страница 170: ...22 Recovering Data 170 ...
Страница 174: ...23 Care and Cleaning 174 ...
Страница 188: ...25 Accessories and Supplies 188 ...
Страница 216: ...26 Specifications and Standards Compliance 216 ...
Страница 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...