Patient Alarms and INOPs
This chapter lists alarms and technical alarms (INOPs) for the fetal monitors, alphabetically,
irrespective of their priority. INOPs start “Technical Alarm Messages (INOPs)” on page 76.
Alarm Messages
Fetal alarms are identified by either "FHR" or "DFHR". All other alarms without these identifiers refer
to maternal parameters.
Alarm Message
From
Condition
Indication
*** BRADY (Pulse)
***BRADY xxx < yyy
SpO
2
The heart rate from the Pulse signal
has fallen below the bradycardia
limit. xxx denotes the lowest
measured value; yyy is the
bradycardia limit.
Numeric flashes and alarm
limit is highlighted, red alarm
message, alarm tone.
*** DESAT
*** DESAT xx < yy
SpO
2
The SpO
2
value has fallen below the
desaturation alarm limit. xx denotes
the lowest measured value, and yy is
the desaturation limit.
Numeric flashes, red alarm
message, alarm tone.
** DFHR1 HIGH
** DFHR2 HIGH
** DFHR3 HIGH
**DFHR1 xxx>yyy
**DFHR2 xxx>yyy
**DFHR3 xxx>yyy
FHR (DECG)
The fetal heart rate obtained from
DECG has risen above the high
alarm limit.
xxx denotes the highest measured
value, and yyy is the high alarm limit.
Numeric flashes and high
limit is highlighted, yellow
alarm message, alarm tone.
** DFHR1 LOW
** DFHR2 LOW
** DFHR3 LOW
**DFHR1 xxx<yyy
**DFHR2 xxx<yyy
**DFHR3 xxx<yyy
FHR (DECG)
The fetal heart rate obtained from
DECG has fallen below the low
alarm limit.
xxx denotes the lowest measured
value, and yyy is the low alarm limit.
Numeric flashes and low limit
is highlighted, yellow alarm
message, alarm tone.
Содержание Avalon FM20
Страница 2: ......
Страница 8: ...8 ...
Страница 56: ...3 Basic Operation 56 ...
Страница 72: ...5 Alarms 72 ...
Страница 82: ...6 Patient Alarms and INOPs 82 ...
Страница 86: ...7 Admitting and Discharging 86 ...
Страница 90: ...8 Non Stress Test Timer 90 ...
Страница 120: ...11 Monitoring Twin FHRs 120 ...
Страница 126: ...12 Monitoring Triple FHRs 126 ...
Страница 136: ...15 Monitoring Uterine Activity Internally 136 ...
Страница 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Страница 154: ...18 Monitoring SpO2 154 ...
Страница 164: ...20 Printing the ECG Waveform 164 ...
Страница 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Страница 170: ...22 Recovering Data 170 ...
Страница 174: ...23 Care and Cleaning 174 ...
Страница 188: ...25 Accessories and Supplies 188 ...
Страница 216: ...26 Specifications and Standards Compliance 216 ...
Страница 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...