Specifications 51
Philips Respironics E30 User Manual
7.1.10 Displayed parameter accuracy
Parameter
Accuracy
Resolution
Range
Estimated Leak Rate
± (5 + 15% of reading) LPM
1 LPM
0 to 200 LPM
Tidal Volume*
± (25 + 15% of reading) mL
1 mL
0 to 2000 mL
Respiratory Rate
greater of ± 1 BPM or ± 10% of reading
1 BPM
0 to 60 BPM
Minute Ventilation*
± (1 + 15% of reading) LPM
1 LPM
0 to 99 LPM
* Accuracies apply with no/low O
2
bleed (less than 10 LPM) in the patient circuit. If higher than, use the correction volume
graph listed below. No tidal volume correction is required when using the inlet O
2
port.
Tidal Volume Adjustment for O2 Flow
Rate in Patient Circuit
0 lpm
10 lpm
20 lpm
25 lpm
30 lpm
A
d
justed
fo
r
O
2
F
lo
w
R
a
te
(
m
L)
Device Reading (mL)
200
300
400
500
600
700
800
900
1000
1100
200
400
600
800
1000
1200
1400
1600
7.2 End of use
The Philips Respironics E30 Ventilator is provided globally for use under local emergency use
authorizations, such as the US Food and Drug Administration (FDA) Emergency Use Authorization (EUA) for
ventilators, Health Canada Interim Order for use in relation to COVID-19, and waiver of CE marking, which
authorize its use for the duration of the COVID-19 public health emergency, unless terminated or revoked
(after which the products may no longer be used).
7.3 Disposal
Separate collection for electrical and electronic equipment per EC Directive 2012/19/EU. Dispose of this
device in accordance with local regulations.
Содержание 1146562
Страница 2: ... 2020 Koninklijke Philips N V All rights reserved ...
Страница 18: ...12 Introduction Philips Respironics E30 User Manual ...
Страница 20: ...14 Therapy modes Philips Respironics E30 User Manual ...
Страница 54: ...48 Cleaning Philips Respironics E30 User Manual ...
Страница 60: ...54 Specifications Philips Respironics E30 User Manual ...
Страница 63: ...Limited warranty 57 Philips Respironics E30 User Manual ...
Страница 64: ...58 Limited warranty Philips Respironics E30 User Manual ...