Device setup 23
Philips Respironics E30 User Manual
Icon
Text
Description
CPAP
S
S/T
PC
Mode
Allows you to select a therapy mode. The default setting is S/T. The available
modes are: CPAP, S, S/T, or PC
Note: If the therapy mode is changed while the blower is turned on, a confirmation
pop-up screen appears. Select
Yes
if you would like to activate the selected mode.
cmH
2
O
cmH
2
O
This setting is available only when CPAP mode is selected. Allows you to increase
or decrease the CPAP pressure setting from 4 to 20 cmH
2
0 in increments of 0.5.
IPAP
IPAP
Increase or decrease the Inspiratory Positive Airway Pressure (IPAP) from 4 to 30
cmH
2
O in increments of 0.5. You cannot set the IPAP setting lower than the EPAP
setting.
EPAP
EPAP
Allows you to increase or decrease the Expiratory Positive Airway Pressure (EPAP)
from 4 to 25 cmH
2
O in increments of 0.5.
BPM
BPM
Allows you to modify the Breaths Per Minute setting. You can choose between Off
and 1 through 30 BPM.
Ti
Ti
Allows you to modify the Inspiratory Time setting. You may adjust the setting from
0.5 to 3.0 seconds in 0.1 increments. Only displays in PC mode and if BPM is not
set to Off in S/T mode.
Alarm settings
Choosing this screen will take you to a sub-menu where you can enable or disable the alarms
described below.
Icon
Text
Description
Patient Disconnect
Alarm
You can enable or disable the Patient Disconnect alarm by
choosing 0, 15, or 60 seconds. The alarm will sound when a large,
continuous air leak is detected in the circuit for more than the
specified alarm setting. The default is 15 seconds.
Apnea Alarm
The Apnea alarm detects the cessation of spontaneous
breathing. You can enable or disable the Apnea alarm by
choosing 0 (off), 10, 20, 30, or 40 seconds. The alarm will sound
when the time between patient-triggered breaths is greater or
equal to the specified apnea alarm setting. The default is 0.
Low Min Vent Alarm
You can enable or disable this alarm by choosing 0 (off) to 99
lpm in 1.0 increments. The alarm will sound when the patient’s
minute ventilation is less than or equal to the specified alarm
setting. The default is 0.
Содержание 1146562
Страница 2: ... 2020 Koninklijke Philips N V All rights reserved ...
Страница 18: ...12 Introduction Philips Respironics E30 User Manual ...
Страница 20: ...14 Therapy modes Philips Respironics E30 User Manual ...
Страница 54: ...48 Cleaning Philips Respironics E30 User Manual ...
Страница 60: ...54 Specifications Philips Respironics E30 User Manual ...
Страница 63: ...Limited warranty 57 Philips Respironics E30 User Manual ...
Страница 64: ...58 Limited warranty Philips Respironics E30 User Manual ...