2-1 

CONTROLS AND CONNECTIONS

 

 
 

FRONT PANEL CONTROLS AND METERS

 

 

Raise Voltage: 

 

Clockwise rotation of this control increases the tester

s output voltage.  The tester

s output  

range is 0-6,000 volts DC.  This control must be returned to the 

0

 (full counter clockwise 

position) before a test can be initiated if either the 

Power

 or 

High Voltage

 switches 

have been previously disengaged, or if an overcurrent condition has been encountered. 

 

Power:

 

When depressed and locked, this illuminated pushbutton switch (green lens) supplies AC 
power to all tester components with the exception of the high voltage power supply. 

 

High Voltage:

 

When depressed and locked, this illuminated pushbutton switch (red lens) supplies AC 
power to the tester

s high voltage power supply. In normal operation this switch will 

illuminate when depressed if the 

Raise Voltage

 control has been set to 

0

.  Illumination 

of this switch indicates that high voltage output is available. 

 

Currentmeter-Megohmmeter:

 

The currentmeter range is selected by the five position rotary switch.  The switch setting 
indicates the full scale current of the currentmeter and the megohmmeter multiplier.  The 
table below shows the full scale currentmeter range with each switch setting. 

 

Currentmeter Ranges 

Switch Setting 

Currentmeter Ranges 

.5 microamps 

0-.5 microamps 

5 microamps 

0-5 microamps 

50 microamps 

0-50 microamps 

500 microamps 

0-500 microamps 

5000 microamps 

0-5000 microamps 

 

The megohm multiplier is selected by the same five position switch as the current-meter 
range.  This allows direct megohm readings at 2.5 and 5 kV.  In order to read megohms 
directly, the kilovolt meter must be set to either 2.5 or 5 kV.  The megohms are then read 
from the right hand meter, megohms at 5 kV on the top scale and megohms at 2.5 kV on 
the middle scale.  The table below shows the megohm range for each switch setting and 
kilovolt range. 

 

Megohmmeter Ranges 

Switch Setting 

2.5 kV 

5 kV 

x10K 

5,000 - 500,000 megohms 

10,000 - 1,000,000 megohms 

x1K 

500 - 50,000 megohms 

1,000 - 100,000 megohms 

x100 

50 - 5,000 megohms 

100 - 10,000 megohms 

x10 

5 - 500 megohms 

10 - 1,000 megohms 

x1 

.5 - 50 megohms 

1 - 100 megohms 

Содержание PM6

Страница 1: ...OPERATIONS AND APPLICATIONS MANUAL MODEL NUMBER PM6 6 KV DC HIGH POTENTIAL TESTER WITH MEGOHMMETER Version 1 0...

Страница 2: ......

Страница 3: ...N NOTES DESCRIPTION SPECIFICATIONS CONTROLS AND CONNECTIONS 2 FRONT PANEL CONTROLS AND METERS INPUT OUTPUT CONNECTIONS INSTALLATION AND OPERATION INSTRUCTIONS 3 INSTALLATION OPERATION TEST LEAD CONNEC...

Страница 4: ...e Test Equipment have been listed below for your reference Become familiar with your instrument before performing an actual test Know your work area check that all circuits are de energized and locked...

Страница 5: ...to keep non essential personnel away from the test site 7 When testing cables the conductors at the distant end should be isolated from each other and taped 8 When testing a faulted cable it is advisa...

Страница 6: ...f motors cables and transformers Input 120 Volts AC 50 60 Hz single phase 3 Amps or 220 Volts AC 50 60 Hz single phase 15 Amps Output 0 6 kV DC continuously adjustable with manual control Negative Out...

Страница 7: ...meter The currentmeter range is selected by the five position rotary switch The switch setting indicates the full scale current of the currentmeter and the megohmmeter multiplier The table below shows...

Страница 8: ...highest level full clockwise approximately 110 of full scale current will be allowed At its lowest level full counterclockwise approximately 10 of full scale current will be allowed before shutdown I...

Страница 9: ...IONS continued AC Line This cord supplies input power to the tester Output This test lead with clip and insulation boot supplies the testers output to the apparatus under test Fuse If this protective...

Страница 10: ...pressing the Power switch The switch will illuminate and remain recessed 2 Energize the High Voltage power supply by pressing the High Voltage power switch The switch will illuminate and remain recess...

Страница 11: ...reading must be multiplied by 4 Similarly if a test voltage of 1 kV is used on the 5 kV scale the corresponding upper megohm scale reading and range multiplier result must be multiplied by 2 NOTE Be...

Страница 12: ...meter 3 Any current flowing through the test lead attached to the Guard terminal on the tester will never be measured by the DC Microammeter 4 Current flowing through the test lead attached to the Gro...

Страница 13: ...nal to the Test Set depending on the design Nothing is connected to the Guard Both stray leakage to earth ground and leakage through the Test Specimen is indicated on the current meter This configurat...

Страница 14: ...ed on the current meter Any current associated with the Guard bypasses the current meter Leakage current through the test specimen is indicated on the current meter Stray leakage to earth ground is no...

Страница 15: ...4 1 SCHEMATIC Drawing Number Description 01 P124 PM6...

Страница 16: ......

Страница 17: ...0 LABEL HV ON 1 PROTECTIVE GROMMET 1 SW3 SW4 SWITCH ROTARY 2 POLE 2 1863047 KNOB FOR ROTARY SWITCH 2 1355310 R23 POT 10K SENSITIVITY ADJ 1 1761092 KNOB FOR ABOVE POT 1355305 ASSY NO 7209900500 SECTION...

Страница 18: ...1116901 SECTION 02 PCB1169 PM6 MAIN PCB 1169 1 1111690 T1 TRANSFORMER ST 4 24 24V CT 120V PRI 1 1894335 K1 RELAY 12VDC COIL 1 1701035 K2 RELAY 120VAC COIL 4PDT 1 1701300 SOCKET 1 1157340 WIRE BALE 1 1...

Страница 19: ...1 1712405 R23 RES 1 5M 1 4W 1 1 1735140 R10 11 33 39 RES 1K 1 1 4W MF 4 1722082 R15 100 OHM 1 4W 1 2 1720180 R19 RES 14 5K 5W 1 1 1722200 R24 22K 1 4W 1 MF 1 1722587 R25 RES 110K 1 4W 1 MF 1 1734110 R...

Страница 20: ...g parts to include the serial number of the unit as well as the part number of the replacement part When your purchase order is received at our office a representative of Phenix Technologies will cont...

Страница 21: ...for Return Cause of Defect If Phenix Technologies Inc deems return of the part appropriate it will then issue an Authorization for Return If return is not deemed advisable other inspection arrangement...

Отзывы: