INSTALLATION
7. A location with enough space to ensure proper ventilation
Allow for a
1/4
" (6.35 mm) space on left side and at the top. Leave 2 ½” (6.35 cm) minimum clearance on
the right side (hinge side) to allow for the door to swing open.
Installation
1. Remove the packaging materials and tapes.
Remove all transportation packaging materials and
tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a diluted
neutral dishwashing detergent (Undiluted detergent can damage the plastic components. For the
dilution, refer to the instruction of the detergent). After cleaning with the diluted detergent, always wipe it
off with a wet cloth, then wipe the panels with a dry cloth.
2. Door handle installation.
Remove the two plugs from the door where you will attach the door handle.
Apply door handle, plastic washer and screw and fasten the screw. For floor-level installation the handle
should be mounted at the higher location. For stacking above another compatible unit, the handle should be
mounted at the lower location.
3. Ground (earth)
WARNING
Use a power supply outlet with ground (earth)
to prevent electric shock. If the power supply outlet is not
grounded, qualified personnel must install a ground.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod.
Such grounding
may cause electric shock in the case of an incomplete circuit.
4. Access Port:
If a temperature sensor is to be placed inside the cabinet, the cable
can be led through the access port on the right side of the cabinet. After using the port, a
rubber cap and insulation should be reinstalled to seal the access port. Make a cut on the
rubber cap and pass the sensor wire through it. (Fig. 1) Failure to do this can affect the
temperature uniformity inside the cabinet and lead to condensation on the outside of the
access port.