INSTALLATION 

 

    Installation site

 

To  operate  this  unit  properly  and  to  obtain  maximum  performance, 

install the unit in a location with the following conditions:

 

1.  A location not subjected to direct sunlight 

Installation in a location subjected to direct sunlight might lead to inadequate cooling. 

2.  A location with adequate ventilation 

Always keep in front of air intake and exhaust area open for free air circulation. Poor ventilation will result 

in inadequate cooling. 

3.  A location away from heat generating sources 

Avoid  installing  the  unit  near  heat-emitting  appliances  such  as  gas  ranges  or  stoves.  Heat  can  reduce 

refrigeration efficiency. 

4.  A location with a sturdy and level floor 

Install the unit on a sturdy floor to avoid vibration and noise. Placing the unit on an unsteady floor may cause 

vibration and noise. 

  WARNING

 

Be sure to install the unit on a sturdy floor.

    If the floor is not strong enough or the installation site is not 

adequate, this may result in injury from the unit falling or tipping over.   

  CAUTION

 

Select a level and sturdy floor for installation.

 This precaution will prevent the unit from tipping. Improper 

installation may result in water spillage or injury from the unit tipping over. 

 

5.  A location not prone to high humidity 

  WARNING

 

Do not use the unit outdoors.

 Current leakage or electric shock may result if the unit is exposed to rain 

water.

 

Never  install  the  unit  in  a  humid  place  or  a  place  where  it  is  likely  to  be  splashed  by  water.

   

Deterioration of the insulation may result which could cause current leakage or electric shock. 

Do not install the unit under water pipes or steam pipes.

 Deterioration of the insulation may result which 

could cause current leakage or electric shock. 

 

6.  A location without a flammable or corrosive gas 

Avoid placing the unit where chemicals are stored or gases are produced. Also avoid areas where there is 

a great deal of dust. 

  WARNING

 

Never install the unit in a flammable or volatile location.

 This may cause explosion or fire.

 

Never install the unit where acid or corrosive gases are present

 as current leakage or electric shock may 

result due to corrosion. 

 

Содержание PF-L5181W-PA

Страница 1: ...ead these instructions carefully before using this product and save this manual for future use See page 22 for Model No PF L5181W PA Laboratory Freezer Operating Instructions PF L5181W PA Left Door Fa...

Страница 2: ...ONS P 17 DEFROST P 17 MAINTENANCE P 18 STACKING TWO UNITS P 20 TROUBLESHOOTING P 21 SPECIFICATIONS P 22 PERFORMANCE P 23 SERVICE P 24 SAFETY CHECK SHEET P 25 1 No part of this manual may be reproduced...

Страница 3: ...y other person Precautions are illustrated in the following way WARNING Failure to observe WARNING signs could result in a hazard to personnel possibly resulting in serious injury or death CAUTION Fai...

Страница 4: ...om the power supply plug before inserting in a power source A dusty plug or improper insertion may pose a hazard Use a power supply outlet with ground earth to prevent electric shock If the power supp...

Страница 5: ...ess and that doors cannot be closed completely The disposal of the unit should be accomplished by appropriate personnel Risk of child entrapment Before you throw away your old freezer take off the doo...

Страница 6: ...r This may cause injury if the unit tips over Always disconnect the power supply plug before moving the unit Take care not to damage the power cord A damaged cord may cause electric shock or fire Disp...

Страница 7: ...te closing or damage of door gasket 8 In the freezer compartment put stored items on the shelves and do not let them contact the wall 9 If an instrument requiring a power source is to be placed inside...

Страница 8: ...use For intended use on laboratories 2 Altitude up to 2000 m 3 Ambient temperature 20oC to 24oC 4 Maximum relative humidity 75 for temperature up to 24oC decreasing linearly to 50 5 Main supply volta...

Страница 9: ...L PRODUCT PF L5181W PA Solid Door Width up to hinge cover 23 27 32 606mm Height up to hinge cover 34 5 32 867mm Depth up to Solid Door 23 21 32 600mm Door Height 30 5 8 777mm Note Power Cord 60 152 4c...

Страница 10: ...ture sensor inside the cabinet for the purpose of external monitoring 2 Door Lock Turn the key clockwise through 180 degrees to lock the door 3 Control Panel The operation status is displayed on this...

Страница 11: ...ghout entire chamber 2 Powder coated wire shelves Evaporator position is fixed and include wires to support the product 3 Magnetic door gasket Magnetic force ensures the door is sealed to provide opti...

Страница 12: ...wn Arrow Pressing this down arrow key in the setting mode causes the numerical value to go down in 1 C increments A long press of this key will enter offset mode Offset mode is used to calibrate the t...

Страница 13: ...p once The display will show 20 on LED Screen and it will start blinking 5 Press key Within 2 seconds of pressing the button you will hear a beep once The display will show 19 on LED Screen and it wil...

Страница 14: ...y result in injury from the unit falling or tipping over CAUTION Select a level and sturdy floor for installation This precaution will prevent the unit from tipping Improper installation may result in...

Страница 15: ...handle plastic washer and screw and fasten the screw For floor level installation the handle should be mounted at the higher location For stacking above another compatible unit the handle should be mo...

Страница 16: ...the freezer temperature is 18oC Note The unit is set at the factory at 20oC Table 2 Basic operation procedure Example of setting Freezer 19oC Operation Key operated Display after the key operation 1 C...

Страница 17: ...play flash 2 Set to 19 by using Up arrow key Pressing the key shifts up the figure by 1 C Repeat until flashing digit turns to 1 3 Press the set key SET The current adjusted temperature offset by 1 de...

Страница 18: ...l sensor goes open E10 or short circuit E1S Either E10 or E1S is displayed Defrost sensor abnormality If the defrost sensor has a short circuit E2S Note Defrost sensor is present in the unit but the d...

Страница 19: ...dishwashing detergent Undiluted detergent can damage the plastic components For the dilution refer to the instruction of the detergent After the cleaning with the diluted detergent always wipe it off...

Страница 20: ...emove the base grille to access the condenser coils for cleaning Procedure to remove base grille 1 Open the freezer door 2 Using a Philips screwdriver remove the two screws 3 Remove the base grille 4...

Страница 21: ...frigerator or SR L6111W PA laboratory refrigerator or SF L6111W PA laboratory freezer to double the floor space efficiency To stack two units you must order the stacking accessory kit part number 833...

Страница 22: ...TURE COLLECTS ON THE OUTSIDE 1 Frequent door openings in highly humid conditions can cause this Reduce door openings and reduce time door is open 2 Make sure door seal is tight FREEZER IS NOISY 1 Make...

Страница 23: ...ht 2 2 lbs 1 kg Maximum load capacity for each shelf 33 lbs 15 kg Access port Inner diameter 1 3 16 30 mm 1 port on the right side Cooling method Forced air circulation Compressor Reciprocal type outp...

Страница 24: ...range 20 C 5 C Rated voltage 115 V AC Rated frequency 60 Hz Power consumption 106 kWh per year These specifications were determined by following testing standard NSF International Standard NSF ANSI 4...

Страница 25: ...24 English SERVICE Wiring Diagram...

Страница 26: ...2 Contamination of the unit Unit interior No contamination Yes No Decontaminated Yes No Contaminated Yes No Others 3 Instructions for safe repair maintenance of the unit a The unit is safe to work on...

Страница 27: ...26 English MEMO...

Страница 28: ...27 English Printed in Mexico ARAKUA100351 1300 Michael Drive Suite A Wood Dale Illinois 60191 USA PHC Corporation of North America...

Отзывы: