background image

51 

 

       

 

Replacement of glow starter

 

 
The glow starter is located at the switch box on the lower right side.     
 

1.

 Turn off the power switch and disconnect the power supply plug from the outlet. 

2.

 Turn the glow starter to counter clockwise to remove. (See figure below) 

3.

 Set a new glow starter. (type: FG-1P) 

• 

The glow starter is consumables. Please replace the glow starter at the same time with the lamp. 

 
 
 
 
 
 
 
 
 
 

Replacement of glow starters 

(Optional light add on kit MIR-L15)

 

 
The glow starters of optional light add on kit (MIR-L15) are located inside the unit cover on the lower left 
side.  
   

1.

 Turn off the power switch and disconnect the power supply plug from the outlet. 

2.

 Remove

 

4 screws fixing the unit cover by using a screw driver. 

3.

 Turn the glow starter to counter clockwise to remove. (See figure below)

 

4.

 Set new glow starters. (type: FG-1P)

 

• 

The glow starter is consumables. Please replace three glow starters at the same time with the lamps. 

 
 
 
 
 
 
 
 
 
 
 
 
 

Note:

    Consult our sales representative or agent when replacing the glow starters. 

 

Glow starter (FG-1P)

 

Glow starter 

(FG-1P)

 

4 screws (fixing the unit cover)

 

Содержание MIR-554 Series

Страница 1: ...54 Series Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 61 for model number Operating Instructions Cooled Indub...

Страница 2: ......

Страница 3: ...SAVED PROGRAM MENU Edit 28 START OF PROGRAM MENU Run 29 JOIN FUNCTION 31 SKIP OF STEP MENU Skip 32 STOP OF PROGRAM MENU Stop 33 AUTOMATIC DEFROST MENU Tools Date Time 34 MANUAL DEFROST MENU M def 35 V...

Страница 4: ...2 CONTENTS TROUBLESHOOTING 52 DISPOSAL OF UNIT 53 LOW HUMIDITY MODE 58 PERFORMANCE DATA 59 SPECIFICATIONS 60 PERFORMANCE 61 SAFETY CHECK SHEET 62...

Страница 5: ...t our sales representative or agent if any page of the operating instructions is lost or the page order is incorrect Contact our sales representative or agent if any point in the operating instruction...

Страница 6: ...ersonnel possibly resulting in serious injury or death CAUTION Failure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property Symbol shows This sym...

Страница 7: ...d safe disposal options visit ec gc ca mercure mercury Dispose of or recycle in accordance with applicable laws Pour plus de renseignements sur les proc dures de manutention s curitaire les mesures pr...

Страница 8: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Страница 9: ...fire When removing the power supply plug from the power supply outlet grip the power supply plug not the cord Pulling the power supply cord may result in electric shock or fire by short circuit Discon...

Страница 10: ...se may cause corrosion of inner components or electric parts Consult our sales representative or agent when you use a material resulting in the corrosive gas such as in the case of insect breeding Che...

Страница 11: ...C Maximum relative humidity 80 for temperature up to 31o C decreasing linearly to 50 relative humidity at 40o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltages...

Страница 12: ...10 INCUBATOR COMPONENTS 1 2 3 4 5 6 7 8 9 10 12 13 14 11 17 18 16 15 19...

Страница 13: ...s opening 10 Access port This port allows cables to be passed into the cabinet Note Be sure to set the rubber stopping to the access port placed at the left and right side as it was when passing cable...

Страница 14: ...key CE To clear the input value during editing of program 9 Enter key ENTER To determine the selection of menu In program editing pressing this key causes moving to a next article 10 Shift key Upward...

Страница 15: ...issible contact capacity DC 30 V 2 A a Contact output Connect the lead wires to COM and NO terminals Normal condition open Abnormal condition close b Contact output Connect the lead wires to COM and N...

Страница 16: ...e a malfunction It is due to the hot gas piped around the unit frame to prevent condensation around the cabinet A location with a sturdy and level floor Always install the unit on a sturdy and level f...

Страница 17: ...an damage the plastic components For the dilution refer to the instruction of the detergent After the cleaning with the diluted detergent always wipe it off with a wet cloth Then wipe off the panels w...

Страница 18: ...4 Display of current value Current value of temperature is displayed 5 Message display field A message is displayed when a breakdown occurs 6 Alarm display Alarm Alarm is alternately displayed by reve...

Страница 19: ...t the running for low humidity page 40 Default setting To set the default for LCD panel and communication DAQ speed etc page 42 STANDBY OPERATION MENU Std by This product automatically operates with s...

Страница 20: ...tting of key unlock Key Unlock 1 When setting of key unlock change the value of the key lock line from 1 to 0 in the stand by setting screen Stand by Setting and press the enter key ENTER 2 Input pass...

Страница 21: ...window is displayed 7 Select in MENU window pressing the shift key downward is displayed in white letters 8 Press the enter key ENTER ENTER Chamber temperature setting value 10 0 is displayed 9 Press...

Страница 22: ...morized and top screen is displayed Procedures of Key Unlock Changing Key Lock setting from 1 Lock to 0 Unlock Operation Key LCD Display 1 Top screen is displayed 2 Press the menu button MENU MENU MEN...

Страница 23: ...5 o C or lower than the minimum temperature in a program or stand by operation 5 Press the menu button MENU at the completion of setting The menu window is shown Select OK and press the enter key ENT...

Страница 24: ...ect Edit and push the enter key ENTER 3 The Program Name Stored screen is opened Press the menu button MENU and select New and press the enter key ENTER The program names are displayed when some progr...

Страница 25: ...hanged by Insert or Delete on the menu window Press the menu button MENU to open the menu window The menu window for a first section has no Insert or Delete Therefore neither insert nor delete is effe...

Страница 26: ...e lighting is not usable The lamp is off with the temperature out of range when the saved program is running In this case temperature fluctuation may exceed 1 5 o C or it may take longer time to pull...

Страница 27: ...e Program screen is opened 1 0St p 31Cyc LS Time 6 00 9 00 11 00 13 00 15 00 17 00 Li ght 0 0 0 0 0 0 LS Time 19 00 21 00 22 00 23 00 Li ght 0 0 0 0 When setting Light step LS displayed at the upper l...

Страница 28: ...saved Edit function of characters Shift key Upward shift key Space insertion Downward shift key backspace Leftward shift key Move a cursor left Rightward shift key Move a cursor right Character input...

Страница 29: ...s displayed and 99 is changed into Display of changes 99 when lapped by a cursor 2 Set a time temperature and light step as same as Example 1 3 Input a program name NIKKO pres the menu button MENU to...

Страница 30: ...menu window Select Save when saving by overwriting or select SaveAs when saving with another program name Press the enter key ENTER The edited program is entered Do not select SaveAs with same program...

Страница 31: ...t OK and press the enter key ENTER Timer selection of Clock mode or Timer mode 1 Clock Clock mode Displays start time of each steps 2 Dec Timer mode Displays the remaining time up to a next step Join...

Страница 32: ...utton MENU to show the menu window Select Graph and press the enter key ENTER When the starting time of program operation is later Waiting is displayed in the current running status of LCD panel O z e...

Страница 33: ...ut NIKKO as the same character string the input for the join function is as follows NIKKO 1 Autumn NIKKO 2 Spring NIKKO 3 Summer 2 When running the joined program in Ex 1 select the program Oze 1 Spri...

Страница 34: ...menu button MENU Selecting Next causes the skip to the next step Selecting Back causes the skip to the previous step After selecting Next or Back press the enter key ENTER 4 Programmed running shifts...

Страница 35: ...and press the enter key ENTER 3 The Confirmation screen is displayed Select OK and press the enter key ENTER to stop the program 4 After stopping the program the top screen is displayed C o n f i r m...

Страница 36: ...displayed on running status of LCD panel During defrosting the chamber temperature deviates from the set temperature The chamber temperature raise by approximately 2 o C with the setting temperature...

Страница 37: ...ing Defrosting is displayed at the upper left on the top screen CAUTION The manual defrost can be started during programmed running standby operation or automatic defrosting 4 The manual defrost is fi...

Страница 38: ...ep The displayed date is scrolled by pressing the leftward or rightward shift key leftward shift key older date rightward shift key newer date S t a n d b y T o p S c r e e n T e mp 2 5 0 o C A l a r...

Страница 39: ...10 o C Similarly select Range1 and press the enter key ENTER Upper limit is changed from 60 o C to 50 o C and lower limit is changed from 10 o C to 0 o C 5 0 o C 2 0 1 0 0 1 0 1 0 0 o C 1 2 2 4 M E N...

Страница 40: ...ation of hyper terminal on PC Apply txt or csy to retention file as an extension Press the menu button MENU to show the menu window Select Start and press the enter key ENTER The transmission is start...

Страница 41: ...d 1 Log interval 2 min About 5 days 2 Log interval 6 min About 14 days 3 Log interval 30 min About 70 days After passing the memory limit the older data is deleted and newer data is memorized Setting...

Страница 42: ...0 4 When optional light add on kit MIR L15 is installed set the lamp Lamp to 1 When optional light add on kit MIR L15 is installed except when the set value of temperature is between 2 o C 50 o C the...

Страница 43: ...ENTER 4 Input User Password 4 digits again and press the enter key ENTER Note Consult our sales representative or agent if you put off the password K e y L o c k P W S e t t i n g C u r r e n t U s e...

Страница 44: ...f temperature in chamber 0 0 o C 50 o C 1 20 o C 80 o C Note DAQ is an external monitoring system of chamber status It is necessary to set the DAQ Speed DAQ ID and DAQ Mode when using communication so...

Страница 45: ...button MENU to show the menu window Select OK and press the enter key ENTER The program Oze 1 Spring is now deleted D e l e t e U s e r D a t a D e l e t e a U s e r P r o g r a m D e l e t e A l l U...

Страница 46: ...ENTER 6 The Confirmation screen is displayed The menu window is already displayed Select OK and press the enter key ENTER All of the programs are now deleted C o n f i r m a t i o n A r e y o u s u r...

Страница 47: ...mber Ensure that the intake and exhaust vents are not blocked Blocking these vents cause deviation of chamber temperature from the set temperature Always install the stainless plate on the bottom shel...

Страница 48: ...heater and sub heater OFF Thermal sensor abnormality Input voltage is over 70 o C corresponding Input voltage is lower than 50 o C corresponding Top screen E01 Temp sensor is opened E02 Temp sensor i...

Страница 49: ...to 50 o C or less disconnection of sensor Top screen E09 Comp sensor is opened Intermittent tone Alarm status Compressor OFF Defrost sensor abnormality The input voltage corresponds to 50 o C or less...

Страница 50: ...if under the program operation The alarm is activated as follows under program operation The buzzer is stopped by pressing the alarm buzzer stop key BUZZER but the remote alarm keeps alarm status The...

Страница 51: ...detergent When a diluted neutral dishwashing detergent is used wipe the cabinet or accessories thoroughly with a cloth soaked in clean water Then wipe the cabinet or accessories unit with a dry cloth...

Страница 52: ...tray on the floor after touching to the floor 4 To replace the evaporating tray follow the procedure in reverse Note Consult our sales representative or agent when cleaning the evaporating tray Replac...

Страница 53: ...dd on kit MIR L15 The glow starters of optional light add on kit MIR L15 are located inside the unit cover on the lower left side 1 Turn off the power switch and disconnect the power supply plug from...

Страница 54: ...erature of chamber reaches to the set temperature of operation The excessive heat source is in the chamber Remove the excessive heat source For the allowable heat load refer to the graphs on page 59 W...

Страница 55: ...ended period ensure that children do not have access and doors cannot be closed completely The disposal of the unit should be accomplished by appropriate personnel Always remove doors to prevent accid...

Страница 56: ...the chemical involved German Entsorgung von Altger ten und Batterien Nur f r die Europ ische Union und L nder mit Recyclingsystemen Dieses Symbol auf den Produkten der Verpackung und oder den Begleitd...

Страница 57: ...piles pictogramme du bas Si ce pictogramme est combin avec un symbole chimique il r pond galement aux exigences pos es par la Directive relative au produit chimique concern Spanish Eliminaci n de Apa...

Страница 58: ...Neste caso estar em conformidade com o estabelecido na Diretiva referente aos produtos qu micos em causa Italian Smaltimento di vecchie apparecchiature e batterie usate Solo per Unione Europea e Nazi...

Страница 59: ...aand symbool Dit symbool kan in combinatie met een chemisch symbool gebruikt worden In dit geval volstaan de eisen die zijn vastgesteld in de richtlijnen van de desbetreffende chemische stof Swedish A...

Страница 60: ...dity mode is selected Do not select the low humidity mode when the drying of culture media should be avoided When selecting the normal mode no low humidity mode Pay attention to the following when the...

Страница 61: ...0 20 30 40 50 60 70 1 0 1 2 3 4 5 6 7 8 9 our Heat load input in chamber Ambient temperature Temperature reached in chamber 0o C Temperature reached in chamber 10o C Maximum heat load tolerable in cha...

Страница 62: ...rocated compressor Output 250 W Evaporator Fin and tube type Condenser Wire and tube type Refrigerant R 404A Defrost heater 322 W Temperature controller Microprocessor PID control compressor ON OFF Te...

Страница 63: ...z 50 Hz Power consumption 290 W 295 W 285 W Environmental conditions Ambient temperature 5 o C 35 o C Humidity less than 80 R H 1 The control range when the optional inner door MIR 55ID installed is 1...

Страница 64: ...f the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see be...

Страница 65: ......

Страница 66: ......

Страница 67: ......

Страница 68: ...Printed in Japan LDCL043300 1 S0418 10418 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018...

Отзывы: