18
REMOTE ALARM TERMINAL
WARNING
Always disconnect the power supply cord before connecting an alarm device to the remote alarm
terminal.
The remote alarm terminal is installed at the back of the unit. The alarm is outputted from this terminal.
Contact capacity is DC 30 V, 2 A.
Contact output:
Contact output:
Terminal
Normal status
Abnormal status
Between COM. and N.O.
open
close
Between COM. and N.C.
close
open
Note:
• The remote alarm is stopped by pressing buzzer stop key (BUZZER) as the remote alarm is operated in
conjunction with the buzzer, except for the power failure alarm.
• The remote alarm terminal is in alarm status when the power supply plug is unplugged because it is
regarded as a power failure.