25
BASIC PARAMETERS
4.
Input each parameter. Press Apply key to
save the input value and setup. The display
returns to the Tools screen.
●Each setting
• Compressor Delay:
The time from turning ON this unit until starting its compressor. This unit is required a large electric power
at the moment its compressor starts. When some units are in a same room, set so as to be shifted to
each other the compressor delay times, to prevent the simultaneous start of all compressors after power
failure. Settable range: 3 minutes~15 minutes, factory setting: 3 minutes.
5.
Press Top key to return to the Top screen.
Setting key lock
1.
Press Menu key to lead the Menu screen.
2.
Press Key Lock key to lead the Key Lock
screen.