background image

 

Instructions For Use: 
 
MCO230AICUV - IncuCyte  kit 

 

1.

 

Follow the installation instruction from the original instructions for use of the MCO-230AIC-PA. 

2.

 

Take out the original humidifying pan, humidifying pan cover and the 4 trays (see Fig. 1 and 2). 

 

 

 
 

3.

 

Place the smaller humidifying pan 
with part number 624 230 4548 on 
the bottom of the incubator and 
make sure to place it with the long 
side against the back wall of the 
incubator so water accumulated on 
the dew stick will drip in this 
humidifying pan. 
 
 
 

4.

 

Place the complete humidifying pan cover with part number 624 
230 4555 over the humidifying tray and make sure to place it 
against the back wall of the incubator.

 

 

 

 
 
 
 

5.

 

Place the stand, NU-STAND-230AICGEN, for the IncuCyte S3 system on the bottom of the incubator 

6.

 

Install the IncuCyte S3 inside the incubator. Set temperature to 36.5 °C (see manual of MCO-230AIC-
PA to adjust).  
 

Installation 

 

Отзывы: