Phase Technology 70Xi Скачать руководство пользователя страница 3

Phase Technology 70Xi-APA/ APC/ ASL Analyzer System – Quick Installation Guide

 

Page 3 of 3 

Customer Service & Support:   604-241-9568 

 

 

[email protected] 

 

18.

 

Power up the analyzer:  

 

Press the round button at the bottom right hand side at the 
front of the analyzer.  
 

 

 

The analyzer will perform a set of self tests to verify 
its main components, including the optics system 
and cooling system, are operating correctly.  

If a problem is detected, the self tests will stop and a 
message will be displayed on the bottom of the 
screen.   

If any component fails the self tests, press the 
INTELLINOSTICS button for further 
information and troubleshooting steps.  

If the pour point function is available, the analyzer 
will automatically perform an auto-purge process 
after the self tests. 

 
19.

 

Power up the autosampler:  

 

Switch the power control above the power cable connector 
to the “I” position.  

 

 

 

The autosampler will perform a series of self-tests, 
moving its arm to pre-defined locations.  

 

 

Warning: The arm moves and rotates at a 
moderately fast speed during its self-tests
The tip of the autosampler probe is sharp. 

 

 

 

Caution: Do not place obstacles that would 
block or impede the movement of the arm. If 
the arm movement is impeded at any point 
during the self-test causing it to stop or 
stutter, switch off the autosampler and 
power it on again to reinitialize. 

 

 

20.

 

After the self-tests are completed, the analyzer shows 
the Vial Setup screen.   

 

 

 

The analyzer returns to this screen whenever 
the VIAL SETUP button is pressed. 

The Sample Cup Temperature, Optics Light Signal, 
and Run Count are displayed at the top of the 
screen. 

A set of Favorites buttons is listed on the right hand 
side of the screen. 

The main area in the middle of the screen is used for 
programming test settings for any vial position.  The 
rows of circles represent the physical vial positions 
available on the two vial racks of the autosampler. 

The ACCESS button on the lower left hand 
corner is used for changing user access levels, 
resetting the analyzer, and shutdown. 

 
 
21.

 

Pressurize the solvent tank: 

 

Press “Analyzer Options” on the Toolbar at the 
bottom of the analyzer screen. 

 

On the Analyzer Options screen, press “Pressurize 
Solvent Tank”. 

 

On the Pressurize Solvent Tank screen, press “Start 
Pressurizing Solvent Tank”. 

 

The pressurization will automatically stop once the 
target pressure is reached, or if the tank failed to 
pressurize (due to the lid not closed properly, or the 
handle of the ball-valve is still in vertical position). 

 

Once the pressurization has successfully completed, 
press “Close” and the analyzer is now ready for 
testing. 

 
 

22.

 

To start a test: 
a)

 

Fill a vial with sample to ¾ full.  Insert a septa 
inside the vial cap (Teflon side facing the sample), 
and screw the vial cap onto the vial.  Place the vial 
into any available position on the sample rack.

 

 

b)

 

To program a vial using a 
predefined test setting, 
simply drag a suitable 
favorite button to the vial 
position where the 
corresponding test sample 
has been placed on the 
sample rack. 

c)

 

Repeat the drag & drop 
procedure if more than one vial needs to 
be programmed.  After all vials have been 
programmed, press the START TEST 
button to start the test sequence. 

Отзывы: