Phase Technology 70Xi-APA/ APC/ ASL Analyzer System – Quick Installation Guide

 

Page 2 of 3 

Customer Service & Support:   604-241-9568 

 

 

[email protected] 

 

 

 
7.

 

Use wrench to connect the tubing to the autosampler: 

Port at autosampler: 

Tubing: 

Elbow fitting at bottom of 

wash pot of vial heater 

 

Labeled  

“Bottom of Wash Pot” 

Vent port at front of wash pot 

of vial heater 

 

½” OD tubing without 

label (may have already 

attached to the vent port) 

 

8.

 

Use wrench to connect the tubing to solvent tank: 

Port label at solvent tank: 

Tubing label: 

“Air In” 

“Air In” 

“Solvent Out” 

 “Solvent Out” 

“Vent” 

“To Solvent Tank Vent” 

 
9.

 

Remove the cap of the waste tank, and carefully screw in the 
pour sprout (with level sensor attached). 

 
10.

 

Use wrench to connect the tubing to the waste tank: 

Port at waste tank: 

Tubing: 

 “From Wash Pot To Waste Tank” 

“To Waste Tank” 

Any of the 

available ports 

“To Waste Tank Vent” 

 
11.

 

For the 2 vent tubes from the solvent tank and waste tank, 
direct the open end to a well-ventilated area such as an 
exhaust fume-hood. Exhaust liquid/fumes can emerge

 

from 

these ports. 

 
12.

 

Connect the vial heater block vent tube to the waste tank. If 
this is not possible, direct the vial heater block vent tube to 
the exhaust. Install a cap on the waste tank barb fitting if 
directing the vial heater block vent tube to the exhaust. 

 
 
13.

 

Connect the sensor/heater cables to the back of the analyzer: 

Port label at back of analyzer: 

Cable label: 

“Transfer Tube Heater” 

“To Transfer Tubing Heater” 

“Level Sensor” 

“To Level Sensor” 

“Heater Block 1” 

“Heater Block 1” 

“Heater Block 2” 

“Heater Block 2” 

 
 
 
14.

 

Connect the sensor cables to the solvent and waste tank: 

Tank: Cable 

label: 

Solvent: “Level Sensor” port 

with cable labeled “S” 

“S” 

Waste: Cable labeled “W” 

“W” 

 
 
 
 

15.

 

Connect the RS232 cable from the “AUTOSAMPLER” port on 
the back of the analyzer to the “HOST” port of the 
autosampler. 

 

 

“AUTOSAMPLER” port on 

back of the analyzer 

 

 

“HOST” port of 

the autosampler 

 
 

16.

 

Connect the AC power cables: 
 

a)

 

Connect the AC power 
cable to the back of the 
analyzer 

 

 b)

 

Connect the AC power 
cable to the back of 
the autosampler 

 

 
Plug the power cables into a grounded AC outlet that is a 
dedicated circuit.  If the power source is unstable or noisy, 
use a power surge protector, equipped with a line filter. 
 
 

17.

 

Fill solvent tank with Heptane:  

a)

 

Lift the handle on the solvent 
tank lid to unlock the lid. 

b)

 

Remove the lid from the solvent 
tank. 

c)

 

Fill the solvent tank with Heptane. 

d)

 

Replace the lid and push the 
handle down to lock it.  Before 
locking it, ensure that there is 
equal spacing around the entire 
lid between the lid and the tank 
for proper sealing. 

e)

 

Push the ball-valve handle down 
to the horizontal position. 