Phase Technology 70Xi-APA/ APC/ ASL Analyzer System – Quick Installation Guide
Page 2 of 3
Customer Service & Support: 604-241-9568
7.
Use wrench to connect the tubing to the autosampler:
Port at autosampler:
Tubing:
Elbow fitting at bottom of
wash pot of vial heater
Labeled
“Bottom of Wash Pot”
Vent port at front of wash pot
of vial heater
½” OD tubing without
label (may have already
attached to the vent port)
8.
Use wrench to connect the tubing to solvent tank:
Port label at solvent tank:
Tubing label:
“Air In”
“Air In”
“Solvent Out”
“Solvent Out”
“Vent”
“To Solvent Tank Vent”
9.
Remove the cap of the waste tank, and carefully screw in the
pour sprout (with level sensor attached).
10.
Use wrench to connect the tubing to the waste tank:
Port at waste tank:
Tubing:
“From Wash Pot To Waste Tank”
“To Waste Tank”
Any of the
available ports
“To Waste Tank Vent”
11.
For the 2 vent tubes from the solvent tank and waste tank,
direct the open end to a well-ventilated area such as an
exhaust fume-hood. Exhaust liquid/fumes can emerge
from
these ports.
12.
Connect the vial heater block vent tube to the waste tank. If
this is not possible, direct the vial heater block vent tube to
the exhaust. Install a cap on the waste tank barb fitting if
directing the vial heater block vent tube to the exhaust.
13.
Connect the sensor/heater cables to the back of the analyzer:
Port label at back of analyzer:
Cable label:
“Transfer Tube Heater”
“To Transfer Tubing Heater”
“Level Sensor”
“To Level Sensor”
“Heater Block 1”
“Heater Block 1”
“Heater Block 2”
“Heater Block 2”
14.
Connect the sensor cables to the solvent and waste tank:
Tank: Cable
label:
Solvent: “Level Sensor” port
with cable labeled “S”
“S”
Waste: Cable labeled “W”
“W”
15.
Connect the RS232 cable from the “AUTOSAMPLER” port on
the back of the analyzer to the “HOST” port of the
autosampler.
“AUTOSAMPLER” port on
back of the analyzer
“HOST” port of
the autosampler
16.
Connect the AC power cables:
a)
Connect the AC power
cable to the back of the
analyzer
b)
Connect the AC power
cable to the back of
the autosampler
Plug the power cables into a grounded AC outlet that is a
dedicated circuit. If the power source is unstable or noisy,
use a power surge protector, equipped with a line filter.
17.
Fill solvent tank with Heptane:
a)
Lift the handle on the solvent
tank lid to unlock the lid.
b)
Remove the lid from the solvent
tank.
c)
Fill the solvent tank with Heptane.
d)
Replace the lid and push the
handle down to lock it. Before
locking it, ensure that there is
equal spacing around the entire
lid between the lid and the tank
for proper sealing.
e)
Push the ball-valve handle down
to the horizontal position.