phase II PHT-2000 Скачать руководство пользователя страница 6

 

 

6

3.3  Notation of Leeb’s Hardness 

 

When measuring the hardness of a sample material using the traditional static 

hardness testing method, a change of applied pressure will result in a change in 

the hardness reading.  This will also happen during a Leeb’s Hardness test when 

you change the impact device.  In measuring the hardness of the same test 

sample with different impact devices, the Leeb’s hardness values obtained will 

vary. 

 

For Example:  720 HLD

720 HLC 

 

 

Because different converting curves are obtained from different impact devices, 

when converting hardness L into another hardness value the notation for the 

converted hardness value should include the impact device used. 

 

For example: 

Hardness HV converted from hardness L using impact device D+15 

Should be written as 22, 8 HV LD+15 

Where:  

22=hardness value HL 

 

8=hardness value HV 

 L=Leeb’s 

Method 

LD+15-impact device 

 

Hardness HRC converted from hardness L using impact device D should be 

written as 35, 9 HRCLD 

Where:  

35=hardness value HL 

 

9=hardness value HRC 

 L=Leeb’s 

Method 

 D=Impact 

device 

 

 

 

4. 

Pretreatment of Samples 

 

4.1  Pretreatment of Sample Surface 

To eliminate hardness errors resulting from the roughness of a sample’s surface when 

using impact device D, DC or D+15, the test surface should be polished until its 
roughness 

Ra

 is no more then 2

µ

m. PHASE II 

SRG-1000

 handheld surface roughness 

tester or any other suitable instrument may be used to measure the surface roughness 

of the sample material.  The test surface should be clean and free from oil stains. 

 

4.2 Curved 

Surface 

When testing a curved surface with a radius of curvature 

R

 less than 1.2 in (30mm), a 

special small support ring washer should be used to compensate the curvature. 

 

 

 

 

 

 

Содержание PHT-2000

Страница 1: ...struction Manual 201 933 6300 www phase2plus com Precision Measuring Tools Indexing Fixtures Workholding Fixtures Hardness Testers Ultrasonic Thickness Surface Finish Coating Thickness Model No PHT 20...

Страница 2: ...production line Obtaining original records of test data 2 2 Technical Data Display error 0 8 at LD 900 Value display Value actual x 100 Value Actual Relative repetitive display error 0 8 at LD 900 Val...

Страница 3: ...ength of 50mm It is suitable for testing in extremely confined spaces the base of grooves and on special components such as gear wheels For use on steel and cast steel only G Probe This device has an...

Страница 4: ...40 173 Bronze 60 290 Copper 45 315 Table 3 Impact Device D 15 Measuring Range LD 15 300 900 Material HRC HRB HB HV HSD Steel Cast St 19 3 67 9 80 638 80 937 33 3 99 3 Table 4 Impact Device C Measuring...

Страница 5: ...nst the sample surface by a spring force and then rebounds back At a distance of 1mm from the sample surface the impact and rebound velocity of the impact body are measured by the following method A p...

Страница 6: ...15 Should be written as 22 8 HV LD 15 Where 22 hardness value HL 8 hardness value HV L Leeb s Method LD 15 impact device Hardness HRC converted from hardness L using impact device D should be written...

Страница 7: ...nimum depth of hardened layer should be 0 03 in 0 8 mm The testing surface of the sample should be perpendicular to the impact direction Note The above restrictions referring to the rigidity of test s...

Страница 8: ...button for 2 seconds 6 2 Setting of all Parameters Choosing an Impact Device Press the button to go to the main menu Use the button to scroll down to 6 D DC Press the button to scroll through the imp...

Страница 9: ...ld the button a second time for 1 second to bring the backlight up to the second brightest stage Press and hold the button for a third time to aquire the brightest back light To turn off the back ligh...

Страница 10: ...s will appear in memory 0001 and will continue to save test results in progressive order There is no need to change any settings Memory Recall While in the main test screen press the button to go to m...

Страница 11: ...s with the axis of the impact device 6 4 Additional Features Alteration of parameters including hardness value and type of sample material can be performed during any measurement set The output result...

Страница 12: ...ct body and support ring Release the impact spring after each test Never apply lubrication of any kind to impact body 7 2 Repair When the tester is calibrated against the Rockwell standard test block...

Страница 13: ...0 0 0 Corrections to be added to Rockwell B F and G values Hardness Diameters of Convex Cylindrical Surfaces Reading 1 4 in 3 8 in 1 2 in 5 8 in 3 4 in 7 8 in 1 in 6 4 mm 10mm 13mm 16 mm 19 mm 22 mm 2...

Страница 14: ...5 8 51 4 471 495 443 229 46 73 6 60 83 5 64 8 50 3 458 480 432 221 45 73 1 59 2 83 0 64 0 49 0 446 466 421 215 44 72 5 58 5 82 5 63 1 47 8 434 452 409 208 43 72 0 57 7 82 0 62 2 46 7 423 438 400 201 4...

Страница 15: ...6 162 81 140 83 51 1 87 6 71 8 55 8 159 173 159 80 137 82 50 6 87 3 71 1 54 8 156 170 156 76 135 81 50 86 9 70 4 53 8 153 167 153 73 133 80 49 5 86 6 69 7 52 8 150 164 150 72 130 79 48 9 86 3 69 1 51...

Страница 16: ...1 78 5 72 3 40 3 8 6 93 72 35 28 7 78 0 72 0 39 6 7 6 92 71 34 28 2 77 0 71 7 39 0 6 6 91 70 33 27 8 76 6 71 4 38 3 5 6 90 69 32 27 4 76 0 71 0 37 6 4 6 89 69 31 27 0 75 5 70 7 37 0 3 6 88 68 30 26 6...

Страница 17: ...72 75 72 71 0 018 0 46 84 66 68 68 64 62 0 020 0 51 82 57 63 55 53 0 022 0 56 69 79 47 58 45 43 0 024 0 61 67 76 94 51 34 31 0 026 0 66 65 71 87 37 18 0 028 0 71 62 67 80 20 4 0 030 0 76 57 60 71 0 03...

Отзывы: