7
Maintenance and Repair
Avoid placing the PHT-2000 near locations where it can be exposed to shock, heavy
dust, dampness, oils or
strong magnetic fields.
7.1
Impact Device
•
Using the included brush, clean the guide tube and impact body after it has
performed 1000-2000 tests using the following procedure:
Screw off the support ring and take out the impact body.
Clean the tube by winding the brush counter-clockwise into the tube until it
reaches the bottom and then pull it out. Repeat this procedure 5-6 times.
Replace the impact body and support ring.
Release the impact spring after each test.
Never apply lubrication of any kind to impact body!
7.2
Repair
•
When the tester is calibrated against the Rockwell standard test block and issues
a reading over 1 HRC, it
could
mean the carbide ball is worn out and should be
replaced.
•
Upon coming across any functional problems not mentioned in this manual,
contact a Phase II Customer Service Representative at (201)933-6300.
•
DO NOT
open the
PHT-2000
or dismount any of its fixed parts. This
precision instrument should only be serviced by a factory trained service
representative.
•
If the unit is opened, its warranty will automatically become void
.