Malfunctions
13
Fig. 8:
Clean transmitter
Detach the transmitter from the vacuum apparatus.
Fill carefully with industrial alcohol; allow to soak for 5 minutes.
Pour out the alcohol and dispose of according to the local regulations.
Allow the measurement chamber to dry sufficiently (at least 10 minutes).
Connect the transmitter to the vacuum chamber.
Evacuate the transmitter.
Bring transmitter up to operating temperature (
≥
5 minutes).
Adjust the transmitter.
8
Malfunctions
8.1
Rectifying malfunctions
industrial
alcohol
Problem
Possible causes
Remedy
Measuring signal continually < 0.5 V;
Display controller: "ErLo" or
"Sensor Error"
●
No supply voltage
●
Transmitter maladjusted
●
Transmitter defective
Check connecting cable,
power supply or controller
Adjust the transmitter
Replace the transmitter
Reading is too high at low pressures
●
Strong outgassing of the sensor
●
Sensor contaminated
Clean the transmitter
Adjust the transmitter
Replace the transmitter