6

Installation

4

Installation

4.1 Electrical connection

Unscrew and remove the cover of the relay box.

Install the wiring of the mains power and the backing pump according to the 
technical data and the following illustrations.

Close the relay box.

DANGER

Voltage-bearing elements

Danger to life from electric shock as a result of improper installation.

Electrical connection may be carried out only by trained and authorised electricians.

Ensure the system is adequately earthed.

Establish an adequate fuse protection on customer side (depending on the model).

Fig. 3:

Backing pump relay box, 5 A, internal view

1

Casing / Relay box

1a

Control lead

2

Cable fitting PG 11

3

Cable fitting PG 9

4

Cable fitting PG 7

Fig. 4:

Backing pump relay box, 20 A, internal view

1

Casing / Relay box

1a

Control lead

2

Cable fitting PG 11

3

Cable fitting PG 9

4

Cable fitting PG 7

L

N

1

3

2

1

a

4

PE

PE

N

L1

i

nsul

a

te!

B

a

ck

i

ng

pump

M

ai

ns

i

nput

n.c. bn

bl

sw

N

2

N

PE

PE

1

L

L1

3

4

1

a

B

a

ck

i

ng pump

M

ai

ns 

i

nput

bl

sw

bn

n.c.

i

nsul

a

te!

Содержание PM 061 372-T

Страница 1: ...Betriebsanleitung Operating Instructions PT 0294 BEN B 1005 Backing pump relay box PM 061 372 T PM 061 373 T PM 061 374 T PM 061 375 T...

Страница 2: ...2 3 Improper use 4 3 Product description 5 3 1 Relay box 1 phase PM 061 372 T PM 061 374 T 5 3 2 Relay box 1 phase PM 061 373 T PM 061 375 T 5 4 Installation 6 4 1 Electrical connection 6 4 2 Connect...

Страница 3: ...on international certification standards as specified by UL CSA ANSI Z 535 SEMI S1 ISO 3864 and DIN 4844 In this document the following hazard levels and information are considered Pictograph defini t...

Страница 4: ...tronic drive unit TC 400 integrated electronic drive unit TC 400 with Y Connector integrated electronic drive unit TC 1200 integrated electronic drive unit TC 1200 with Y Connector integrated electron...

Страница 5: ...vacuum pumps The maximum relay contact load is 5 A 3 2 Relay box 1 phase PM 061 373 T PM 061 375 T The relay box PM 061 373 T PM 061 375 T with mechanical relay is provided for the operation of large...

Страница 6: ...by trained and authorised electricians Ensure the system is adequately earthed Establish an adequate fuse protection on customer side depending on the model Fig 3 Backing pump relay box 5 A internal v...

Страница 7: ...1 PE F2 4 A F1 4 A 3 x 1 5 mm2 Mains input 1 2 4 3 Potential separation 6 mm U1 X3 1 3 2 X1 N L 3 x 1 5 mm2 PE N L Backing pump n c M8 M12 n c PM 061 372 T PM 061 374 T M8 M12 insulate M PE N L1 PE 3...

Страница 8: ...ng pump Make the settings and control via the interfaces of the electronic drive unit TC 400 Connect the control lead of the relay box directly to a vacant accessory port of the electronic drive unit...

Страница 9: ...P 350 Connect the control lead of the relay box to the remote connection of the elec tronic drive unit using an adapter cable Establish the mains supply for the relay box according to the accessory op...

Страница 10: ...th accessory ports and bridges other lengths on re quest PM 061 552 T TCS 12 adapter for TC 110 with interface RS 485 and 4 accessory ports PM 061 638 U Y Connector M12 for accessories P 4723 012 Adap...

Страница 11: ...1 Technical data 6 1 Dimensions Fig 7 PM 061 372 T PM 061 374 T Fig 8 PM 061 373 T PM 061 375 T 73 49 24 57 19 5 65 94 PG 9 PG 7 PG 11 120 85 A B A B 52 25 40 80 35 40 80 18 5 12 5 18 5 PG 9 PG 7 PG 1...

Страница 12: ...ective 2004 108 EEC and EU Low Voltage Directive 2006 95 EEC PM 061 372 T PM 061 373 T PM 061 374 T PM 061 375 T Guidelines harmonised standards and national standards and specifications which have be...

Страница 13: ...0 64 41 8 02 2 02 info pfeiffer vacuum de www pfeiffer vacuum net Vacuum is nothing but everything to us Turbopumps Rotary vane pumps Roots pumps Dry compressing pumps Leak detectors Valves Component...

Отзывы: