Installation
– Depending on the pump type, use metallic hoses or PVC hoses with flange
connections.
– Separators, filters etc. may be installed upstream to protect the pump (see ac-
cessories). However, please observe the loss of pumping capacity due to the
conductivity of the accessories.
5.3 Connecting the exhaust side
Choose the cross-section of the exhaust line to be at least the size of the nominal
connection diameter of the vacuum pump's exhaust connection.
Piping to the pump must be suspended or supported.
– Physical forces from the piping system must not be allowed to act on vacuum
pumps.
Lay piping from the pump sloping downward so that no condensate can flow
back into the pump; otherwise fit a condensate separator.
– If an air trap is created in the system, then a device for draining condensation
water must be provided at the lowest point.
5.4 Connecting to the mains power supply
The input voltage may lie in the ranges 100 ... 120 VAC (50/60 Hz) and 200 ... 240
VAC (50/60 Hz). The pump is preset ex works to a voltage range of 200 ... 240 VAC.
In case of another supply voltage the input has to be switched over accordingly.
CAUTION
High pressure in the exhaust line!
Danger of damage to the seals and danger of the pump bursting.
Install the line without shut-off valves on the exhaust side.
If there is danger of a build-up of excess pressure (> 1500 mbar abs.) in the lines,
observe all official accident prevention safety regulations.
If the exhaust gases are being extracted, the exhaust pressure must be at least
250 mbar greater than the pressure at the intake side.
WARNING
Emission of toxic substances from the exhaust!
Danger of poisoning from emitted gases or vapours, which can be detrimental to
health and/or can pollute the environment, depending on the particular application.
Comply with the applicable regulations when working with toxic substances.
Only officially approved filter systems may be used to separate and remove these
substances.
DANGER
Voltage-bearing elements
Danger to life from electric shock.
The electrical connection can be carried out only by trained and authorised electrici-
ans.
Ensure the system is adequately earthed.