10

Operation

6

Operation

NOTICE

Damage to the pump due to condensate in the operating fluid!

Returned condensate generated by vapors or by temperature differences between the 
oil mist filter and pump impairs the quality of the operating fluid and negatively impacts 
the pump's final vacuum.

Drain operating fluid built up with condensate in a time manner.

NOTICE

Risk of the filter elements becoming blocked with resin!

When pumping gases and vapors with a tendency towards polymerization, the ONF fil-
ter elements can become resinified.

Observe the corresponding safety devices such as the saturation indicator or pres-
sure relief valve.

NOTICE

Make sure that the return for the operating fluid works.

Operating fluid is only sucked in and returned from an operating pressure of < 100 hPa 
and starting from a minimum quantity of operating fluid in the ONF.

Long evacuation phases with a high intake pressure should always be followed by 
operating phases with a lower operating pressure.

Содержание ONF 16 M

Страница 1: ...Translation of the original instructions Operating Instructions EN PK 0213 BEN L 1412 ONF 16 M 25 M ONF 25 L XL Oil Mist Filter...

Страница 2: ...Storage 5 4 Product description 5 4 1 Product identification 5 4 2 Function 6 5 Installation 7 5 1 Assembly 7 5 2 Installing the operating fluid return line 8 6 Operation 10 7 Maintenance 11 7 1 Drai...

Страница 3: ...ecified by UL CSA ANSI Z 535 SEMI S1 ISO 3864 and DIN 4844 In this document the following hazard levels and information are considered Pictographs DANGER Imminent danger Indicates an imminent hazardou...

Страница 4: ...t of rotary vane pumps Use the ONF in accordance with the corresponding approved pumping speed 2 3 Improper use Improper use will cause all claims for liability and warranties to be forfeited Improper...

Страница 5: ...uct when communicating with Pfeiffer Vacuum always have the information from the rating plate available Model and model number Date of manufacture Fig 1 Product identification on the rating plate exam...

Страница 6: ...ion becomes excessive so that the maximum working pressure of 1500 hPa absolute cannot be exceeded The collect ed operating fluid can be observed through a sight glass and discharged with a drain scre...

Страница 7: ...the flange and fasten it with the clamping ring accessory or bracket screws accessory if clamp flanges are used observe centering ring accessory Lay exhaust line from the ONF sloping downward so that...

Страница 8: ...O ring 138 Fit operating fluid return hose 222 at both sides Fasten hose clips 223 Fig 3 DUO Rotary vane pump with ONF and operating fluid return line ONF 25 XL Unscrew operating fluid drain screw 33...

Страница 9: ...d remove the sealing plug 78 Fit operating fluid return hose 222 at both sides Fasten hose clip 223 Tighten the union nut of the screw fitting 77 Fig 4 PentaLine pump with ONF and Operating fluid retu...

Страница 10: ...becoming blocked with resin When pumping gases and vapors with a tendency towards polymerization the ONF fil ter elements can become resinified Observe the corresponding safety devices such as the sat...

Страница 11: ...k to the rotary vane pump Dispose of condensate according to the respectively valid legal requirements Operating fluid return line Due to the automatic operating fluid return it is not necessary to dr...

Страница 12: ...2 from the cover 4 Caution The parts are under preload Detach casing 2 and filter insert 8 and all other internally located parts from cover 4 take care with O rings 28 not with ONF 25 XL and 30 Repla...

Страница 13: ...urface of valve buffer 14 replace if necessary When assembling lightly oil valve buffer and ensure the correct seating in the valve casing 6 Assembling Continue assembly by reversing the dismantling s...

Страница 14: ...ed Drain off operating fluid If possible send pump or unit in the original packaging Sending of contaminated pumps or devices No units will be accepted if they are contaminated with micro biological e...

Страница 15: ...to Uno 30 M standard version PK 196 944 T Oil return unit from ONF 25 M 25 L 25 XL to Penta PK 198 545 T Oil return unit from ONF 16 M to Duo 5 M from ONF 25 M to Duo 10 M from ONF 25 L to Duo 20 M st...

Страница 16: ...L Ma e ONF 16 M ONF 25 M ONF 25 L A 142 mm 147 mm 218 mm B 80 mm 80 mm 80 mm C 120 mm 132 mm 134 mm D 127 mm 127 mm 182 mm E 16 mm 16 mm 25 mm F 105 mm 105 mm 105 mm G 17 mm 23 mm 27 mm H 48 mm 48 mm...

Страница 17: ...are parts packages ONF 16 M 25 M ONF 25 L Fig 8 ONF 16 M 25 M ONF 25 L Spare parts package Model No Consisting of Maintenance kit ONF 16 M PK E37 001 T 8 28 30 ONF 25 M PK E37 001 T 8 28 30 PK E37 003...

Страница 18: ...12 Disposal Products or parts thereof mechanical and electrical components operating fluids etc may cause environmental burden Safely dispose of the materials according to the locally applicable regu...

Страница 19: ...are the only supplier of vacuum technology that provides a complete product portfolio Benefit from our know how and our portfolio of training opportunities We support you with your plant layout and pr...