3. Safely disconnect the drive motor from the mains.
4. Secure the motor against switching back on.
5. Remove the vacuum pump from the system if necessary.
6. Dispose of used lubricant according to applicable regulations in each case.
7. For maintenance work, only dismantle the vacuum pump to the extend needed.
8. Only use alcohol or similar media to clean the pump parts.
9. Avoid residues of cleaning agents inside the vacuum pump.
7.2 Checklist for inspection and maintenance
You can carry out maintenance work at
Maintenance Level 1
yourself.
We recommend that Pfeiffer Vacuum Service (PV) carry out maintenance work at
Maintenance Level 3
(inspection).
The interval for
Maintenance Level 3
(overhaul) can be extended if you use vibration monitoring. Com-
ply with the specifications in the "Vibration monitoring" section. Pfeiffer Vacuum recommends an interval
of four years in this case.
If the required intervals listed below are exceeded, or if maintenance work is carried out improperly, no
warranty or liability claims are accepted on the part of Pfeiffer Vacuum. This also applies if original
spare parts are not used.
Action
Inspec-
tion
Mainte-
nance
level 1
Mainte-
nance
level 2
Mainte-
nance
level 3
Required mate-
rial
described in document
OI
OI
Not rel-
evant
SI
Interval
daily
≤ 1 year
≤ 2
years
Inspection
Visual and acoustic pump check
● Checking lubricant level and
color of the lubricant
■
● Checking vacuum pump for
leaks
● Checking vacuum pump for
running noises
■
Maintenance level 1
Clean the vacuum pump
● Pump housing from outside
● Flushing suction chamber with
a suitable cleaning agent
adapted to the process
■
as re-
quired
Lubricant
● Change lubricant and check o-
rings for damage; replace if
needed
■
Maintenance level 3
OI:
Operating instructions,
SI:
Service instructions
Depending on the process, the required maintenance intervals can be shorter than the reference val-
ues specified in the table. Please consult Pfeiffer Vacuum if necessary.
Maintenance
43/62