WARNING
Risk of crushing on rotating parts when reaching into the open flange
The pistons continue to run in the vacuum after switching off the motor, and can trap fingers and
hands within their reach.
► Wait until the vacuum pump comes to a complete standstill.
► Secure the vacuum pump against re-start.
CAUTION
Danger of burns on hot surfaces
Depending on the operating and ambient conditions, the surface temperature of the vacuum pump
can increase to above 70 °C.
► Provide suitable touch protection.
CAUTION
Health hazard from increased noise emission
Remaining in the close proximity of the vacuum pump for a sustained period of time may cause hear-
ing damage.
► Ensure adequate sound insulation.
► Wear hearing protection.
Risks during maintenance, decommissioning and malfunctions
WARNING
Danger of injury from strong magnetic field
There is a risk of injury for people with pacemakers and medical implants.
► Make sure that such individuals do not enter the sphere of influence (≤ 2 m) of the magnetic field.
► Identify rooms in which magnetic couplings are openly accessible with the symbol: "
No access
for people with pacemakers
".
► Always keep disassembled couplings away from computers, data carriers, and other electronic
components.
WARNING
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work,
there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic sub-
stances causes environmental damage.
► Take suitable safety precautions and prevent health hazards or environmental pollution by toxic
process media.
► Decontaminate affected parts before carrying out maintenance work.
► Wear protective equipment.
WARNING
Risk of crushing from rotating parts
Fingers and hands may be caught by rotating pistons within the connection flange. This results in se-
vere injuries.
► Keep limbs out of the reach of the roots pump.
Safety