24 V DC output / pin 7:

 Inputs 2 to 6 are activated if they are connected to pin 7 (active high) with +24 

VDC. They can also be activated via an external SPS. The functions are deactivated by "SPS high lev-
el" and by "SPS low level".

● PLC High level: +13 V to +33 V
● PLC Low level: -33 V to +7 V
● Ri: 7 kΩ
● I

max

 < 200 mA

6.1.2 Inputs

The digital inputs switch various electronic drive unit functions. Inputs are assigned with functions ex-
factory. You can configure them via the RS-485 interface and the Pfeiffer Vacuum parameter set.

DI remote priority/pin 2
V+:

The "remote" connection has control priority over all other control sources.

open:

Remote priority "inactive"

DI1 (rotation speed setting mode)/pin 3
V+:

Rotation speed setting mode "active"

open:

Rotation speed setting mode "inactive"

DI pumping station/pin 5
V+:

Vacuum pump on.
Control of all components connected and malfunction acknowledgement.

open:

Vacuum pump off

DI stand-by / pin 6
V+:

Stand-by activated
The stand-by speed can be selected in the range 

30 to 100%

 of the nominal speed.

GND/

open:

Stand-by off

6.1.3 Outputs

The digital outputs have a maximum load limit of 24 V/50 mA per output. All outputs listed are configura-
ble with the Pfeiffer Vacuum parameter set via the RS-485 interface (description relates to factory set-
tings).

DO1/Pin 8

V+:

No errors
After applying the voltage supply, the digital output DO1 permanently outputs V+ meaning 

which means "no error".

GND

Error
"Active low" signifies "Error" (common error message).

DO2/Pin 9

V+:

Vacuum pump on
"Active high" means ''Vacuum pump on'' and rotates at set rotation speed.
Example: Use the signal for the message vacuum pump "Ready for operation".

GND:

Vacuum pump off

6.1.4 RS-485

Pin 13 and pin 14

You can connect a Pfeiffer Vacuum display and control unit (DCU or HPU) or an external PC via pin 13 
and pin 14 on the D-sub connection of the electronic drive unit.

Interfaces

24/52

Содержание MVP 010-3 DC

Страница 1: ...OPERATING INSTRUCTIONS EN Translation of the Original MVP 010 3 DC Diaphragm Pump...

Страница 2: ...pfeiffer vacuum de Further operating instructions from Pfeiffer Vacuum can be found in the Download Center on our website Disclaimer of liability These operating instructions describe all models and v...

Страница 3: ...on 16 3 1 Function 16 3 1 1 Drive 16 3 1 2 Pumping system 16 3 1 3 Cooling 16 3 2 Connections 17 3 3 Identifying the product 17 3 4 Shipment 17 4 Transportation and Storage 18 4 1 Transporting the vac...

Страница 4: ...m pump 36 9 Maintenance 37 9 1 Maintenance instructions 37 9 2 Checklist for inspection and maintenance 38 9 3 Change the diaphragms and valves 38 9 3 1 Dismantle the diaphragm head and valves 38 9 3...

Страница 5: ...tronic drive unit for diaphragm pumps when deliv ered 32 Tbl 13 Configure parameters P 019 and P 024 33 Tbl 14 Parameters P 060 and P 061 33 Tbl 15 Behavior and meaning of the LED of the electronic dr...

Страница 6: ...Fig 5 Connecting the electronic drive unit to the power supply DCU and turbo pump 22 Fig 6 Pin assignment of the D Sub socket 15 pin 23 Fig 7 Cross link via connection RS 485 using connection cables a...

Страница 7: ...oup These operating instructions are aimed at all persons performing the following activities on the product Transportation Setup Installation Usage and operation Decommissioning Maintenance and clean...

Страница 8: ...tickers on the product 1 2 Fig 1 Position of the stickers on the product 1 Rating plate 2 Hot surface warning sign 1 3 4 Abbreviations Abbreviation Meaning in this document DC Direct current DCU Displ...

Страница 9: ...tion with a short description Example P 312 software version PA Polyamide PE Earthed conductor protective earth T Temperature in C property abbreviation of the vacuum pump TC Turbopump electronic driv...

Страница 10: ...ICE Danger of property damage Notice is used to address practices not related to physical injury Instructions on avoiding property damage Notes tips or examples indicate important information on the p...

Страница 11: ...es being expelled without an exhaust line Exhaust gases and vapors are released from the turbo pumping station unhindered during normal us age In the case of processes with toxic media there is a risk...

Страница 12: ...equirements for double isolation between mains input voltage and output voltage in accordance with IEC 61010 and IEC 60950 Make sure that the power supply pack meets the requirements in accordance wit...

Страница 13: ...ecommended by Pfeiffer Vacuum 2 6 Foreseeable improper use Misuse of the product invalidates all warranty and liability claims Any use that is counter to the purpose of the product whether intentional...

Страница 14: ...troubleshooting and maintenance of the product and who have the following qualifications Qualification in the mechanical field in accordance with nationally applicable regulations Knowledge of this d...

Страница 15: ...level 3 Customer with Pfeiffer Vacuum service training Pfeiffer Vacuum service technician 2 7 3 Advanced training with Pfeiffer Vacuum For optimal and trouble free use of this product Pfeiffer Vacuum...

Страница 16: ...Cover lid 8 Fixing plate 3 Connecting socket 9 Diaphragm head 1 4 LED 10 Vacuum connection G 1 8 5 Diaphragm head 3 11 Exhaust with silencer G 1 8 6 Diaphragm head 2 3 1 1 Drive The diaphragm pump for...

Страница 17: ...Identifying the product To ensure clear identification of the product when communicating with Pfeiffer Vacuum always keep all of the information on the rating plate to hand Learn about certifications...

Страница 18: ...adequately sized level surface Transporting the vacuum pump without its packaging 1 Unpack the vacuum pump 2 To protect the inside of the pump leave the blind plugs on the vacuum connection during tra...

Страница 19: ...m connection lines with large nominal diameter prevents pressure losses Condensate separator Pfeiffer Vacuum recommends the installation of a condensate separator in case vapors are formed from moistu...

Страница 20: ...flange 3 Depending on the pump type use PVC or metallic hoses with flange connections from the Pfeiff er Vacuum component shop 4 Option In case of higher gas throughputs mount an exhaust line 5 Route...

Страница 21: ...ween mains input voltage and output voltage in accordance with IEC 61010 and IEC 60950 Make sure that the power supply pack meets the requirements in accordance with IEC 61010 and IEC 60950 Where poss...

Страница 22: ...t 1 Make sure that you have the correct supply voltage 2 Make sure that the S1 power supply pack master switch is off prior to connection 3 Plug the 15 pin connection cable plug into the remote connec...

Страница 23: ...the D Sub socket 15 pin Pin Function Description factory setting 1 24 VDC input Voltage supply for drive and interface 2 DI access request V Control via DIs GND open Control unlocked 3 DI1 V Rotation...

Страница 24: ...nts connected and malfunction acknowledgement open Vacuum pump off DI stand by pin 6 V Stand by activated The stand by speed can be selected in the range 30 to 100 of the nominal speed GND open Stand...

Страница 25: ...Start bits 1 Stop bits 1 Tbl 6 Features of the RS 485 interface Connecting a Pfeiffer Vacuum display and control units or PC Use the connection cable from the control unit scope of delivery or from t...

Страница 26: ...only ASCII code characters 32 127 the exception being the end character of the telegram C R Basically a master e g a PC sends a telegram which is answered by a slave e g electronic drive unit or tran...

Страница 27: ...3 1 0 3 0 9 0 6 0 0 0 6 3 3 0 3 7 cR ASCII 49 50 51 49 48 51 48 57 48 54 48 48 48 54 51 51 48 51 55 13 6 3 4 Telegram example 2 Control command Switch on the pumping station parameter P 010 device add...

Страница 28: ...corresponds with false 1 corresponds with true 7 u_short_int Positive whole number 03 000 to 999 10 u_expo_new Positive exponential number The last of both digits are the exponent with a deduction of...

Страница 29: ...tion speed P 002 stand by mode on off Three digit number of the parameter Display Display of parameter description Description Brief description of the parameters Functions Function description of the...

Страница 30: ...mands 7 3 Status requests Display Designations Functions Data type Access type Unit min max de fault can be stor ed 303 Error code Error code 4 R 309 ActualSpd Actual rotational speed Hz 1 R Hz 310 Dr...

Страница 31: ...ditional parameter for the DCU Additional parameter in the control panel The basic parameter set is set in the electronic drive unit ex factory For controlling con nected external components e g vacuu...

Страница 32: ...pump to the vacuum equipment Before commissioning make completely sure that the pressure side pressure is below the per missible limit Important settings and function related variables are factory pr...

Страница 33: ...or individual requirements you can configure most connections for the electronic drive unit with the parameter set 8 3 1 Configuring the digital outputs Option Description 1 No error active with troub...

Страница 34: ...iffer Vacuum display and control unit observe the associated operating in structions DCU operating instructions available from the Download Center HPU operating instructions available from the Downloa...

Страница 35: ...le only for operation with DCU or HPU Display Activity Meaning None No adequate power supply 10 active 1 Hz No errors Vacuum pump OFF Vacuum pump stationary 50 active 1 Hz No errors Vacuum pump OFF Va...

Страница 36: ...09 or by pressing the button on the DCU Switch the running vacuum pump off and back on again by pressing the on the DCU once 8 6 Switching off the vacuum pump Procedure 1 Allow the vacuum pump to run...

Страница 37: ...against reactivation Dismantle the vacuum pump for inspection away from the system if necessary NOTICE Damage caused by unsuitable cleaning agents Unsuitable cleaning agents damage pump parts Use only...

Страница 38: ...if maintenance work is carried out improper ly no warranty or liability claims are accepted on the part of Pfeiffer Vacuum This also applies if original spare parts are not used Action Inspection Main...

Страница 39: ...on the intake side 5 Remove the intermediate plate 6 Release the valves and the sealing ring from the intermediate plate 7 Lift the diaphragm slightly at the side 8 Unscrew and remove the diaphragm fr...

Страница 40: ...hragm has a right hand thread Pay attention to the spacer disks 3 Place the valves and the sealing ring in the intermediate plate 4 Position the intermediate plate 5 Put on the diaphragm head cover Ob...

Страница 41: ...with the original sealing plugs 4 Store the vacuum pump in dry dust free rooms within the specified ambient conditions 5 Pack the vacuum pump together with a drying agent in a plastic bag and seal the...

Страница 42: ...an error code that you can read out via the interface of the electronic drive unit If an error occurs the vacuum pump switches off Should malfunctions occur you can find information about potential c...

Страница 43: ...display and control panel or the PC 2 Remove the cause of the malfunction 3 Reset the error message with parameter P 009 or by pressing the button on the DCU Error code Problem Possible causes Remedy...

Страница 44: ...acuum Service Error E042 Hardware error EPROM checksum incorrect Contact Pfeiffer Vacuum Service Error E043 Hardware error E2 PROM write error Contact Pfeiffer Vacuum Service Error E090 Internal devic...

Страница 45: ...and addresses can be found on our website in the Pfeiff er Vacuum Service section Advice on the optimum solution is available from your Pfeiffer Vacuum contact partner For quick and smooth handling of...

Страница 46: ...ontainers e Observe the applicable transport conditions ERKL RUNG KONTAMINIERUNG 6 Affix the declaration of contamination to the outside of the packag ing 7 Then send your product to your local Servic...

Страница 47: ...parts only 3 When ordering the overhaul kit observe the respective part number of the diaphragm pump 1 2 3 Fig 10 Spare parts of the inspection sets 1 Valve plate 3 Diaphragm 2 Sealing ring on the su...

Страница 48: ...ose connection 8 6 mm P 4131 029 G Push in T fitting G 1 8 P 4131 030 E Mains cable 230 V AC CEE 7 7 to C13 3 m P 4564 309 ZA Mains cable 115 V AC NEMA 5 15 to C13 3 m P 4564 309 ZE Mains cable 115 23...

Страница 49: ...sion table Pressure units mbar l s Pa m3 s sccm Torr l s atm cm3 s mbar l s 1 0 1 59 2 0 75 0 987 Pa m3 s 10 1 592 7 5 9 87 sccm 1 69 10 2 1 69 10 3 1 1 27 10 2 1 67 10 2 Torr l s 1 33 0 133 78 9 1 1...

Страница 50: ...ump parts Substances in contact with the media Diaphragms EPDM Valves EPDM Head covers Aluminium Hose connection PVC Elbow union CuZn nickel plated Straight compression coupling on the suction hose Cu...

Страница 51: ...f the use of certain hazardous substances delegated directive 2015 863 EU Harmonized standards and applied national standards and specifications EN ISO 12100 2010 DIN EN 55011 2018 05 EN 1012 2 2011 1...

Страница 52: ...PU0071 ed B Date 1910 P N PU0071BEN...

Отзывы: