3.4. Connecting to Power Supply
Power connections must comply with local
regulations. Permissible input voltage range of
the power supply unit must correspond to the
available mains voltage. The power supply unit
may only be connected to mains current with
earthed conductor.
The pump has integrated reverse voltage
protection. The pump will not start if the polarity
is reversed.
➡
Operate the pump with safety extra-low voltage of 24 V in
accordance with the specifications on the rating plate.
– Supply power with power unit TPS 100 or TPS 150 only.
➡
Connect (earth) pump base to frame potential as defined.
Stand-alone operation with TPS 100
➡
Connect power supply via connection cable (PM 051 421 -T)
with TPS 100.
– The pump starts immediately when the power unit is
switched on; the MVP 006 does not have a separate
power switch.
Operation with turbopump in turbo mode operation
with TPS 150
If the MVP 006-4 is to be operated with a TPD 011 or TMH 071,
Turbo mode must be selected.
➡
Connect power supply via connection cable (PM 061 192 -T)
with TPS 150.
– The pump starts immediately when the TPS power unit
is switched on; the MVP 006 does not have a separate
power switch.
☞
PLEASE NOTE
WARNING
Connecting diagram MVP 006-4 in turbo mode
PM 061 192 -T
TPS 150
(TPS 100 with cable PM 051 421 -T)
MVP 006-4
DCU (option)
TMH 071
TCS 010
TC 100