Typical connection:
In this installation, we use:
A primary isolation valve V1 between the turbo pump and the
roughing pump;
- a secondary isolation valve V2 between the turbo pump and the
chamber to be pumped;
- a purge valve V3;
- a venting valve V4.
We recommend that you fit a separate earth (ground) conductor to earth
pump. Use an un-insulated braid or a separate insulated green/yellow
conductor with a minimum 9AWG (3mm2) conductor. Use a M4 x 8 screw and
a shakeproof washer fitted to the earth hole from the pump to secure the
earth conductor to the pump. The impedance between the pump-body and
the earth connection point must be < 0.1 Ohm at 25 A.