Transport and Storage / Safety
12
2.0 Transport and Storage
Please keep the original packaging for later transport, e.g. for calibra-
tion. Any transport damage due to faulty packaging will be excluded
from warranty claims.
Instruments must be stored in dry and closed areas.
In the case of an instrument being transported in extreme temper-
atures, a recovery time of minimum 2 hours is required prior to
instrument operation.
3.0 Safety
EVOMEX Rotary Field Tester XTROID have been constructed and veri-
fied in compliance with the latest safety standards for Test Instruments
IEC/EN 61010-1and IEC/EN 61557-7 and have left the factory in safe
and perfect conditions.
In order to avoid electrical shock, the valid safety and VDE regula-
tions regarding excessive contact voltages must receive utmost
attention, when working with voltages exceeding 120V (60V) DC
or 50V (25V)rms AC. The values in brackets are valid for limited
ranges (as for example medicine and agriculture).
Prior to usage ensure perfect instrument function.
The safety can no longer be insured if the instrument:
• shows obvious damage
• does not carry out the desired measurements
• has been stored for too long under unfavourable conditions
• has been subjected to mechanical stress during transport.
The instrument may not be opened, divided or changed in any way.
The instrument may be used only with the recommended acces-
sories. The use of unsuitable accessories is inadmissible.
The respective accident prevention regulations established by the
professional associations for electrical systems and equipment
must be strictly met at all times.
Avoid any heating up of the instrument by direct sunlight to ensure
perfect functioning and long instrument life.