11
SPECIFICATIONS
Ranges
Reading
coefficient
Accuracy
Voltage
drop at the
terminals
Protection
A DC
A AC
100
scale
30 scale
DC
AC
100 mV (1)
(1)
(1)
2,5 %
(1)
1A fuse
100 µA
X 1
2 %
180 mV
300 µA
X 10
700 mV
1 mA
X 0,01
900 mV
3 mA
X 0,1
950 mV
10 mA
10 mA
X 0,1
2,5 %
1 V
30 mA
30 mA
X 1
1 V
100 mA
100 mA
X 1
1,05 V
300 mA
300 mA
X 10
1,1 V
1 A
1 A
X 0,01
1,4 V
3 A
3 A
X 0,1
1,2 V
10 A (2)
10 A (2)
X 0,1
2,5 % 3,5 %
1,15 V
10A fuse
(1) The 100mV range is used to measure DC current with external shunts.
Reading coefficient
: depends on the shunt used
Consumption of the range
: 64.5µA
(2) Common switch position for the 3A and 10A ranges. Above 3A, don't use the
socket but the specific 10A input socket (use the COM and 10 A sockets for connection).
For a current greater than 7A, limit the duration of the measurement to 3 minutes.
+