6. Remove the load(s) and [0.0] shall be displayed. Calibration is complete.
In case of inconsistency repeat the process.
Please check availability of calibration weights with specialized stores.
VIII. Elimination of problems and errors
Display
Eliminations of errors
None
Remove isolation strip in battery compartment. Check
polarity of batteries. If still no display replace batteries.
[OUt_2]
Start up and operate the scale on a stable horizontal
surface.
[LO]
Replace batteries. Pay attention to their polarity.
[O-Ld]
Attention overload! Reduce the weight immediately.
Wrong weight
indication
Check the weighing unit selected and change it with the
MODE
key if applicable. For extremely precise weighing
results calibrate the scale regularly or after intense use.
VIIII. Specifications
Weighing units:
kg-g / lb-oz
Maximum capacity:
3000g / 6.6 lb
Division (d):
0.1g / 0.01 oz
Resolution:
1 : 30000
Precision:
± 0.2g
Tare range:
up to 100% of capacity (reduces weighing
range)
Auto switch off:
after 120 seconds with no change of load
Operating Temperatures:
5°
– 35°C / 41° – 95°F
Operating humidity:
Max. 70% relative humidity
User calibration:
Auto calibration with optional 2 x 1000g
weights or 1 x 1000g and 1 x 2000g weights
or 4 x 500g
Power supply:
4 x AA 1.5V alkaline batteries or
DC 6.0V jack (optional AC/DC adapter not
supplied by PESOLA)
Measures:
20.0 x 15.4 x 3.2 cm
Net weight with batteries:
672g / 23.7 oz
Manufacturer warranty:
2 years from date of purchase
Copyright 2015 - 2018