ESP User Manual
Programming & parameter settings
9
Parameter settings
* Recommended range for most users.
Example:
The wheelchair is too fast in speed position 5.
To reduce the maximum speed, go to the programming mode and step to
parameter 9.
Press the “S” button to see the current setting. Adjust the value downwards
using the minus button.
Save the value set by again pressing the “S” button.
Switch off and then start the wheelchair with the main power switch on the
control panel to test-drive the wheelchair.
Remember that the maximum speed is reduced if the seat lift is raised. The seat
lift must, therefore, be in its lowest position during the test drive
.
Number
Standard
Range*
Min. Max. Step
Designation
1
15
10 - 75
5
95
5
Max. acceleration (speed 5)
2
15
5 - 50
5
95
5
Max. acceleration (speed 1)
3
75
20 - 75
5
95
5
Max. deceleration (speed 5)
4
30
10 - 70
5
95
5
Max. deceleration (speed 1)
5
40
25 - 75
5
95
5
Max. turn acceleration (speed 5)
6
15
10 - 50
5
95
5
Min. turn acceleration (speed 1)
7
60
35 - 80
5
95
5
Max. turn deceleration (speed 5)
8
40
30 - 60
5
95
5
Min. turn deceleration (speed 1)
9
95
20 - 95
5
95
5
Max. forward speed (speed 5)
10
15
10 - 80
5
95
5
Min. forward speed (speed 1)
11
25
10 - 50
5
95
5
Max. reverse speed (speed 5)
12
10
10 - 30
5
95
5
Min. reverse speed (speed 1)
13
80
50 - 90
5
95
5
Max. turning speed (speed 5)
14
65
25 - 80
5
95
5
Min. turning speed (speed 1)
15
2
N/A
1
5
1
Attendant high speed selection
16
1
N/A
1
5
1
Attendant low speed selection
Содержание ESP
Страница 1: ...ESP USER MANUAL Enhanced Steering Performance US ...
Страница 2: ......
Страница 4: ...How to contact Permobil Head Office of the Permobil group ...
Страница 12: ...ESP User Manual Notes 12 ...
Страница 13: ...ESP User Manual Notes 13 ...
Страница 14: ......
Страница 15: ......
Страница 16: ...Product code 205008 US 0 ...