ESP User Manual
Resetting and calibrating
11
Resetting the basic settings and calibrating
the sensor
Calibrate the sensor and return to the settings the ESP module had on delivery
from Permobil:
-
if the wheelchair does not keep a sufficiently straight course when driven
forwards.
-
after problems with EPROM data. Error code E6, see page 10.
Resetting and calibrating
1.
The main power switch on the control panel must be off.
2.
Hold down the “S”, plus and minus buttons while starting the wheelchair with
the main power switch on the control panel.
The display displays “CS”. The parameters are reset to the values they had
on delivery from Permobil and sensor calibration begins.
Sensor calibration can take up to ten minutes.
When calibration is complete, “—” is displayed in the display.
If calibration fails, “Er” (error) is displayed in the display.
Calibration can be repeated from point one if the error message “Er” was
displayed.
3.
Exit the programming mode by switching off the wheelchair with the main
power switch on the control panel.
Important!
-
The ESP module must be completely still during the calibration process.
-
Calibration must take place at room temperature, 18 to 25°C.
-
The ESP module should be on for at least 20 minutes before calibration
begins. This is to ensure that the sensor has the correct working temperature.
Содержание ESP
Страница 1: ...ESP USER MANUAL Enhanced Steering Performance US ...
Страница 2: ......
Страница 4: ...How to contact Permobil Head Office of the Permobil group ...
Страница 12: ...ESP User Manual Notes 12 ...
Страница 13: ...ESP User Manual Notes 13 ...
Страница 14: ......
Страница 15: ......
Страница 16: ...Product code 205008 US 0 ...