SPECIFICATIONS
AND
FEATURES
See appendix and refer to drawing PCD-3000-01-01 THRU -04 for details of the different models
SYSTEM
INSTALLATION
6.1
S
YSTEM
E
NCLOSURE
The system enclosure is constructed of fiberglass. It is UL and C UL listed and certified to NEMA 4, 4X standards
(UL 508).
The system is designed to be wall mounted vertically with the inlet at the top.
It is suitable for outdoor installation. If installed outdoors, it is highly recommended to install it in a location
that does not get direct sunlight or provide a method of shading the system. In addition, a method of
protecting the system from direct rain is also highly recommended.
6.2
M
OUNTING
Mount the system enclosure to a wall or other rigid structure using four maximum ¼” (6mm) screws install
ed
through the provided holes. (The holes are concealed when the cover is in place).
The system must be installed on a vertical surface with the heated umbilical line entry located at the top. See
below for mounting screw locations.