6.3.4
Filter Drain Outlet
Vent the exhaust to a remote location, connect a line to the port
labelled “Drain Port”. For runs under 10 ft.,
1/4
”
ID tubing is acceptable. For runs 10-40 ft., 3/8
”
ID tubing is recommended, and the number of elbows
should be kept to a minimum. This port must be connected to an appropriate drain that can accept potentially
acidic condensate and, in the case of model PCD-3000-W, a minimum of 0.25 scfm (7 lpm) of air with entrained
acidic mist. See below for location of the drain port on the specific models.
WARNING!
Failure to properly connect this port may result in personal injury and/or property damage!
Connect the line from the filter drain outlet to designated collection/exhaust basin that is
capable of accepting acid mist and the sample gas present at this location.